What happened
Beijing Lepu Diagnostics Technology built GluRing and GluRing s, rings that pass near-infrared light through a finger to estimate blood glucose without a needle. The report says that an artificial-intelligence model predicts the glucose value and that the apps and server contain a deep learning algorithm. The Center for Medical Device Evaluation (CMDE) of China's National Medical Products Administration reviewed the application for Class III registration. The trial was the maker's own, a prospective study of 96 subjects at three clinical institutions, with venous glucose as the reference. The mean absolute relative difference (MARD) was 9.20% (95% CI 8.65% to 9.74%). Of the readings, 99.32% fell in Clarke error grid zones A and B. MARD rose from 7.22% on day 1 to 10.40% at about day 30. No adverse event was related to the device. In a report posted on 12 August 2026, CMDE concluded that “该产品上市带来的受益大于风险” (the benefits of putting the product on the market outweigh the risks). Users must calibrate the ring by finger-stick at least every 30 days, and its estimates may not guide decisions on medication.
Sources
-
Regulatory document, by authors with no tie to the system.
平均绝对相对误差值(MARD%)结果及 95%CI 为 9.20%(8.65%,9.74%)
经综合评价,在当前认知水平上,认为该产品上市带来的受益大于风险,综合剩余风险可接受。
Details
AI system
- Reported configuration
-
Release 1 of the GluRing apps and server (full version 1.1.0)
- Workflow
-
An LED on one side of the ring sent near-infrared light through the finger to an optical sensor on the other side. An artificial-intelligence model in the apps and server predicted blood glucose from the light signals. Each user first calibrated the ring with finger-stick readings while fasting and two hours after a meal.
Follow-up
- Conditions for replication
-
Users are adults with type 2 diabetes who do not use insulin. Users repeat the finger-stick calibration at least every 30 days and keep testing finger-stick blood. The estimates do not serve to screen for or diagnose diabetes or to decide on medication. The ring is used for self-management outside medical institutions.
Entry details
- Sector
-
Healthcare.
- Purpose
-
Beijing Lepu Diagnostics Technology built GluRing and GluRing s to estimate blood glucose without a needle, for self-management by adults with type 2 diabetes who do not use insulin.
30 more questions not yet checked
model, provider, role, configuration, inputs and outputs, autonomy, human oversight, AI contribution, beneficiaries, reach, distribution and access, duration, every measurement question, every money question, ownership and control, dependencies, cross-border connections, geopolitical implications, replication attempts, availability, institutional decision, grounds, action and result, success criterion, every evidence question, observed or intended, alternative assessments, related entries, other identifiers, category mappings.