Skip to content

Medicare credits the Guardian implant with faster heart attack visits

CMS found that Guardian alarms got heart attack patients to hospital faster. In the trial, median arrival took 51 minutes with alarms on and 30.6 hours with alarms off.

  • 2 for a person

    Trial patients with coronary events reached a medical facility in a median of 51 minutes with alarms on and 30.6 hours with alarms off, though fewer deaths or heart attacks were not statistically shown.

Depth runs from 1, a convenience or time saved, to 5, a life or a livelihood preserved.

Status
Reported benefit.
Evidence

found, the grade of this recordstudiedreportedclaimed

Two independent regulatory documents support the benefit.

Compared with

Implanted ALERTS patients randomized to deactivated alarms. Measured.

Without the system
30.6 hours for control subjects
With it
51 minutes for the treated subjects
The system’s part
The ALERTS trial randomized implanted patients to active or deactivated alarms, so the arrival times isolate the alarms' effect. The expanded analysis that CMS relied on also used follow-up from after the randomized period. CMS records the applicant's statement that "proprietary machine learning algorithms" assess the ST segment. FDA's summary describes a rule that compares each beat's ST shift with a 24-hour running baseline against a programmable threshold. Neither document shows a model trained on data.
Not yet checked
LimitsCosts and harmsLearns from dataLater outcome
Benefit type
Improved Physical Health
System
AngelMed Guardian System
Made by
Angel Medical Systems, Inc.
Used by
Sites of the ALERTS trial in the United States, which the documents do not name
Where
United States
Stage
In a pilot
Date
, when the source was published
Kind
Aggregate result
Broad benefit
Monitoring Processes, Materials, or Surroundings, Identifying Objects, Actions, and Events
Collections
Scrutiny and findings
Last revised
1 November 2026

Sources

  1. Regulatory document, by authors with no tie to the system.

    Centers for Medicare & Medicaid Services. Medicare Program: Hospital Outpatient Prospective Payment and Ambulatory Surgical Center Payment Systems and Quality Reporting Programs (CY 2022 final rule with comment period). Federal Register 86 FR 63458, 16 November 2021 (document 2021-24011), section IV.A.2, AngelMed Guardian System, 86 FR 63583-63589.

    Given the additional endpoints evaluated in the expanded analysis that specifically show faster visits for real events while not increasing unnecessary emergency department visits, we agree that the Guardian[supreg] system meets the substantial clinical improvement criterion
  2. Regulatory document, by authors with no tie to the system.

    U.S. Food and Drug Administration, Center for Devices and Radiological Health. Summary of Safety and Effectiveness Data: AngelMed Guardian System, premarket approval P150009. Approval order 9 April 2018.

    The AngelMed device improves diagnostic accuracy of ACS in a clinically meaningful portion of the intended use population, and the risks associated with use of the device are acceptable.

Details

AI system

Workflow

The implant sensed the heart's electrical signal through a lead in the right ventricle and compared ST-segment changes with the patient's baseline over the prior 24 hours. When it detected an abnormal shift, the implant vibrated and an external device flashed and beeped. Patients were told to seek urgent care when the alarm went off, even without symptoms.

Follow-up

Conditions for replication

The patient has had a prior acute coronary syndrome event and remains at high risk of another. The patient can feel the implant's vibration and respond to the alarms. Patients still act on symptoms when no alarm sounds.

Entry details

Sector

Healthcare (cardiology).

Purpose

Angel Medical Systems built the AngelMed Guardian System to detect acute coronary syndrome events from the heart's electrical signal and alert patients to seek urgent care.

30 more questions not yet checked

model, release, provider, role, configuration, inputs and outputs, autonomy, human oversight, beneficiaries, reach, distribution and access, duration, every measurement question, every money question, ownership and control, dependencies, cross-border connections, geopolitical implications, replication attempts, availability, institutional decision, grounds, action and result, success criterion, every evidence question, observed or intended, alternative assessments, related entries, other identifiers, category mappings.

Cite this record

Peter H. Vartanian. (2026). Medicare credits the Guardian implant with faster heart attack visits (AIBD-1006). Artificial Intelligence Benefits Database, edition of 1 November 2026. https://petervartanian.xyz/aibd/records/AIBD-1006.html (archived at https://doi.org/10.7910/DVN/AADSII, version 7.0)

BibTeX Data Field definitions