What happened
Diabeloop's DBLG1 System links a Dexcom G6 glucose sensor, a Kaleido insulin pump and a locked smartphone whose software drives the pump. Patients confirm mealtime doses. The device committee of France's national health authority (HAS) adopted its opinion on 28 January 2020. The main evidence was the SP7 trial, an open-label randomized crossover trial at 12 French centers from March 2017 to August 2018. Among 63 adults with type 1 diabetes who completed both 12-week periods, time in the 70-180 mg/dL range was 68.5% with DBLG1 against 59.4% with pump and sensor alone. Time below 70 mg/dL fell from 4.3% to 2.0%. HbA1c and patient satisfaction did not differ. The committee rated the system a moderate improvement (ASA III) for adults whose HbA1c stayed at 8% or above on pump therapy. It noted that most trial patients had a lower HbA1c and that the trial used an earlier generation of the system. Trial patients also had 48 hours in hospital and round-the-clock support from a nurse and an engineer. HAS made renewal depend on a study of all users.
Sources
-
Assessment report, by authors with no tie to the system.
La Commission considère que le système DBLG1 est susceptible d'apporter un bénéfice sur le contrôle métabolique des patients diabétiques de type 1 sous pompe à insuline et dont le taux d'HbA1c est ≥ 8% depuis plus de 6 mois. Les données disponibles mettent en évidence un effet favorable sur différents paramètres de suivi instantané de l'équilibre glycémique.
Details
AI system
- Reported configuration
-
DBLG1 algorithm v2017.04.20 in the SP7 trial (listed system: software version 1.3.0 of 16 May 2019)
- Workflow
-
A sensor measured interstitial glucose every 5 minutes and sent it to the smartphone. The algorithm drove the insulin pump automatically, and patients confirmed mealtime doses. Trial patients had 48 hours of training in hospital and round-the-clock support from a nurse and a Diabeloop engineer.
Follow-up
- Conditions for replication
-
Patients are adults with type 1 diabetes on pump therapy who receive specific training and education. Patients use the system in closed-loop mode at least 75% of the time, the threshold that HAS set for continuing.
Entry details
- Sector
-
Healthcare (type 1 diabetes).
- Purpose
-
Diabeloop built the DBLG1 System to adjust insulin delivery automatically from continuous glucose readings for adults with type 1 diabetes.
29 more questions not yet checked
model, provider, role, configuration, inputs and outputs, autonomy, human oversight, beneficiaries, reach, distribution and access, duration, every measurement question, every money question, ownership and control, dependencies, cross-border connections, geopolitical implications, replication attempts, availability, institutional decision, grounds, action and result, success criterion, every evidence question, observed or intended, alternative assessments, related entries, other identifiers, category mappings.