Skip to content

Willem, Cardiologs and ILR-ECG-A cut false heart-monitor alerts

Avalia-t found that AI algorithms re-reading insertable heart monitor recordings removed 60.3% to 98% of false-positive alerts in three studies. It rated the evidence as low in quality.

  • Not rated an organization

    The algorithms re-read stored recordings in diagnostic studies, and Avalia-t found no data on the review time that clinicians saved.

Depth runs from 1, a convenience or time saved, to 5, a life or a livelihood preserved.

Status
Partly reported benefit.
Evidence

found, the grade of this recordstudiedreportedclaimed

An independent assessment report supports the benefit.

Compared with

Alerts from the Medtronic Reveal LINQ monitor without the AI filter (Mittal et al. study). Measured.

Without the system
53,9% (IC 95%: 51.4%-56.5%) positive predictive value
With it
74,5% (IC 95%: 71,8%-77%) with the Cardiologs algorithm
The system’s part
Experts adjudicated the same events that the monitors flagged and the algorithms re-read, so the false alerts removed are the algorithms' own share. Avalia-t included only algorithms that use AI, and its evidence tables describe deep or convolutional neural networks for all three. No study measured the effect on clinicians' work.
Not yet checked
LimitsCosts and harmsLearns from dataLater outcome
Benefit type
Improved Task Accuracy
System
Willem (IDOVEN), Cardiologs and ILR-ECG-A (Implicity), AI algorithms that re-read insertable cardiac monitor recordings
Made by
Idoven (Willem), Cardiologs and Implicity (ILR-ECG-A)
Used by
Remote monitoring centers in the studies that Avalia-t reviewed: Valley Hospital in the United States (Cardiologs), one center in Italy (IDOVEN) and 31 US and French centers on the Implicity platform
Where
Spain (assessment), with studies in the United States, Italy and France
Stage
Tested in a laboratory
Date
, when the source was published
Kind
Aggregate result
Broad benefit
Identifying Objects, Actions, and Events, Monitoring Processes, Materials, or Surroundings
Collections
Scrutiny and findings
Last revised
31 October 2026

Sources

  1. Assessment report, by authors with no tie to the system.

    Unidad de Asesoramiento Científico-técnico, Avalia-t, Agencia Gallega para la Gestión del Conocimiento en Salud (ACIS), for the Red Española de Agencias de Evaluación de Tecnologías Sanitarias y Prestaciones del SNS. López Loureiro I, Puñal-Rioboo J, Mejuto Martí T. Evaluación del rendimiento diagnóstico y la seguridad de la inteligencia artificial en la monitorización cardíaca con dispositivos insertables. Informes de Evaluación de Tecnologías Sanitarias. Madrid: Ministerio de Sanidad, 2025. NIPO 133-25-043-1. Resumen, Conclusiones.

    Los algoritmos de IA muestran un VPP entre el 74,5% y el 89,74%. Así mismo, producen una reducción de FP entre el 60,3% y el 98%.
    AI algorithms demonstrated PPV values between 74,5% and 89,74%, reducing FP rates by 60,3% to 98%.

Details

AI system

Workflow

The algorithms re-read the ECG episodes that insertable monitors had stored as possible arrhythmias. Electrophysiology experts classified each episode as the reference standard.

Follow-up

Conditions for replication

The monitor sends its recordings to a remote platform where the algorithm can analyze them. Each algorithm is checked on the monitors that it reads, since performance varied with the device and its software.

Entry details

Sector

Healthcare (cardiology).

Purpose

Idoven, Cardiologs and Implicity built Willem, the Cardiologs algorithm and ILR-ECG-A to re-read insertable cardiac monitor recordings and screen out false arrhythmia alerts.

30 more questions not yet checked

model, release, provider, role, configuration, inputs and outputs, autonomy, human oversight, beneficiaries, reach, distribution and access, duration, every measurement question, every money question, ownership and control, dependencies, cross-border connections, geopolitical implications, replication attempts, availability, institutional decision, grounds, action and result, success criterion, every evidence question, observed or intended, alternative assessments, related entries, other identifiers, category mappings.

Cite this record

Peter H. Vartanian. (2026). Willem, Cardiologs and ILR-ECG-A cut false heart-monitor alerts (AIBD-0926). Artificial Intelligence Benefits Database, edition of 1 November 2026. https://petervartanian.xyz/aibd/records/AIBD-0926.html (archived at https://doi.org/10.7910/DVN/AADSII, version 7.0)

BibTeX Data Field definitions