What happened
Aillis, Inc. developed nodoca, a throat camera with software that analyzes the image and interview data for signs of influenza. Japan's Pharmaceuticals and Medical Devices Agency (PMDA) dated its review report 15 February 2022. A trial enrolled 708 patients at 11 sites in Japan and tested each with the device, a rapid antigen test and PCR. The earlier device missed its sensitivity goal, so Aillis changed the program and the camera. The approved version was then run on stored data from 659 of the trial patients. Against PCR, it reached 76.0% sensitivity and 88.1% specificity, and PMDA found that the lower confidence limits of 70.7% and 85.5% met the goals set in advance. In the trial, patients rated the pain of the throat photo at 0.8 on average, against 5.1 for the swab of the rapid test. Aillis could not show that the device's sensitivity was non-inferior to that of the rapid test. PMDA therefore judged that it should be approved only as an aid to a doctor's diagnosis and should not replace the rapid test.
Sources
-
Regulatory document, by authors with no tie to the system.
Pharmaceuticals and Medical Devices Agency (PMDA). 審査報告書 nodoca(ノドカ), review report on the application by Aillis, Inc., 15 February 2022, published with the deliberation report of the Ministry of Health, Labour and Welfare dated 9 March 2022. Summary of review results and section 4 (総合評価).
追加試験における感度と特異度の 95%信頼区間の下限値は、それぞれ 70.7%、85.5%であり、事前に設定された性能目標は満たされたため、本品の有効性は示されたと判断した。
検体採取にともなう痛みの度合いや涙流頻度がイムノクロマト法検査と比べ有意に低かったことから、本品の安全性に関して特段の懸念はないと判断した。
Details
AI system
- Workflow
-
The camera photographed the back of the throat, and the software analyzed the image with 26 items of interview data such as temperature and symptoms. The approved version was run on stored data from 659 patients of a trial at 11 sites in Japan, after Aillis changed the program and the camera.
Follow-up
- Conditions for replication
-
Patients are at least 6 years old and have influenza-like symptoms. A doctor makes the diagnosis with other clinical findings, and the device does not replace the rapid antigen test. Other infections that produce lymphoid follicles in the throat may need other tests.
Entry details
- Sector
-
Healthcare.
- Purpose
-
Aillis built nodoca to help doctors diagnose influenza by analyzing photographs of the throat together with interview data.
Not yet checked: model, release, provider, role, configuration, inputs and outputs, autonomy, human oversight, AI contribution, beneficiaries, reach, distribution and access, duration, every measurement question, every money question, ownership and control, dependencies, cross-border connections, geopolitical implications, replication attempts, availability, institutional decision, grounds, action and result, success criterion, every evidence question, observed or intended, alternative assessments, related entries, other identifiers, category mappings.