Avive AED exceeded AHA goals for spotting shockable rhythms

In bench tests that FDA reviewed, the Avive AED's neural network detected coarse ventricular fibrillation with 98.9% sensitivity and normal rhythm with 100% specificity.

  • Not rated a person

    The tests used stored recordings, so no patient in cardiac arrest was treated.

Depth runs from 1, a convenience or time saved, to 5, a life or a livelihood preserved.

Status
Intended benefit.
Evidence

found, the grade of this recordstudiedreportedclaimed

An independent regulatory document supports the benefit.

Compared with

American Heart Association performance goals for arrhythmia analysis. Measured.

Without the system
Above 90% sensitivity for coarse ventricular fibrillation and above 99% specificity for normal sinus rhythm
With it
98.9% sensitivity and 100% specificity
The system’s part
Not isolated. The bench tests measured the detector against labeled recordings and published goals. No comparison of rescues with and without the detector is reported.
Limits
Not yet checked.
Costs and harms
Not yet checked.
Later
Not yet checked.
Benefit type
Improved Task Accuracy
System
Avive AED System rhythm recognition detector
Made by
Avive Solutions, Inc.
Used by
Avive Solutions (bench tests on a held-out set of ECG recordings)
Stage
Tested in a laboratory
Date
, when the source was published
Kind
Aggregate result
Broad benefit
Identifying Objects, Actions, and Events
Collections
Scrutiny and findings
Last revised
31 October 2026

Sources

  1. Regulatory document, by authors with no tie to the system.

    U.S. Food and Drug Administration, Center for Devices and Radiological Health. Summary of Safety and Effectiveness Data (SSED), P210015: Avive Automated External Defibrillator (AED) System. Approved 2022-10-31.

    The bench testing demonstrates that the rhythm recognition detector exceeds both the AHA recommendations and the IEC 60601-2-4 standard for the basic safety and essential performance of cardiac defibrillators, for sensitivity and specificity for detecting shockable and non-shockable arrythmias.

Details

AI system

Workflow

The network was developed on non-overlapping training, validation and test sets from more than 23,000 recordings. Test recordings came from defibrillators with multifunction electrodes, one record per person.

Follow-up

Conditions for replication

The patient is unresponsive and not breathing normally. Rescuers use Child Mode for children under 8 years old or under 55 pounds.

Entry details

Sector

Healthcare.

Purpose

Avive Solutions built the Avive AED to let trained and untrained rescuers treat sudden cardiac arrest, with software that decides whether to deliver a shock.

Not yet checked: model, release, provider, role, configuration, inputs and outputs, autonomy, human oversight, AI contribution, beneficiaries, reach, distribution and access, duration, every measurement question, every money question, ownership and control, dependencies, cross-border connections, geopolitical implications, replication attempts, availability, institutional decision, grounds, action and result, success criterion, every evidence question, observed or intended, alternative assessments, related entries, other identifiers, category mappings.