What happened
BrainScope of Bethesda, Maryland, makes the Ahead 100, a handheld EEG device that records from the forehead. Its algorithm combines EEG features into a classification that was set in a separate earlier study. In the B-AHEAD II study, 817 patients were enrolled at 11 emergency departments in the United States, and 552 were analyzed. Treating physicians did not see the device output, and three neuroradiologists read the CT scans. FDA reviewers report a sensitivity of 78.5% and a specificity of 48.6%. Both fell short of endpoints that required lower confidence limits above 78% and 50%. The negative predictive value was 89.5% at a prevalence of 21%. Using educated assumptions, FDA estimated that about 160 patients, 29%, might have been spared exposure to CT. FDA granted the De Novo request on 17 November 2014. The reviewers note concerns about missing data and about how the need for CT was judged.
Sources
-
Regulatory document, by authors with no tie to the system.
U.S. Food and Drug Administration, Center for Devices and Radiological Health. De Novo Classification Request for BrainScope Ahead 100, Models CV-100 and M-100 (DEN140025): decision summary. Granted 2014-11-17.
While the clinical study failed to meet the predefined primary endpoints, the use of the Ahead 100 device may provide adjunctive information about patients when the clinician does not deem it necessary to order a CT scan, adding confidence in the clinical decision path.
Details
AI system
- Workflow
-
Staff recorded EEG from frontal electrodes in the emergency department within 24 hours of the injury. The classification algorithm was fixed before the validation study.
Follow-up
- Conditions for replication
-
The patient is 18 to 80 years old with a closed head injury and a Glasgow Coma Scale score of 13 to 15. Clinicians use the result with standard assessment and never as a substitute for a CT scan.
Entry details
- Sector
-
Healthcare.
- Purpose
-
BrainScope built the Ahead 100 to help clinicians evaluate adults with mild head injury who are being considered for a head CT.
Not yet checked: model, release, provider, role, configuration, inputs and outputs, autonomy, human oversight, AI contribution, beneficiaries, reach, distribution and access, duration, every measurement question, every money question, ownership and control, dependencies, cross-border connections, geopolitical implications, replication attempts, availability, institutional decision, grounds, action and result, success criterion, every evidence question, observed or intended, alternative assessments, related entries, other identifiers, category mappings.
Cite this record
Peter H. Vartanian. (2026). Ahead 100 EEG test missed its goals but helped rule out injury on CT (AIBD-0786). Artificial Intelligence Benefits Database, edition of 31 October 2026. https://petervartanian.xyz/aibd/records/AIBD-0786.html (archived at https://doi.org/10.7910/DVN/AADSII, version 4.0)