Skip to content

EyeBOX missed its accuracy goals but helped rule out concussion

In FDA's review of EyeBOX, the eye-tracking test missed its sensitivity and specificity goals in 282 patients with head injury, but its negative predictive value was 94.5%.

  • Not rated an organization

    The test met neither accuracy goal, and the study did not measure whether it changed any diagnosis.

Depth runs from 1, a convenience or time saved, to 5, a life or a livelihood preserved.

Status
Reported and intended benefits.
Evidence

found, the grade of this recordstudiedreportedclaimed

An independent regulatory document supports the benefit.

Compared with

Pre-specified goals for the lower confidence limits of sensitivity and specificity. Measured.

Without the system
Lower limits above 70%
With it
Sensitivity 80.4% and specificity 66.1%, with neither goal met
The system’s part
Not isolated. The study compared the score with a clinical reference standard. It did not compare diagnosis or care with and without the device. The decision summary refers to an underlying model applied to the processed gaze data and does not show that the model learned from data.
Limits
Not yet checked.
Costs and harms
Not yet checked.
Later
Not yet checked.
Benefit type
Improved Task Accuracy
System
EyeBOX
Made by
Oculogica, Inc.
Used by
Oculogica (clinical study in 282 patients with head injury)
Stage
In a pilot
Date
, when the source was published
Kind
Aggregate result
Broad benefit
Identifying Objects, Actions, and Events
Collections
Scrutiny and findings
Last revised
31 October 2026

Sources

  1. Regulatory document, by authors with no tie to the system.

    U.S. Food and Drug Administration, Center for Devices and Radiological Health. De Novo Classification Request for EyeBOX (DEN170091): decision summary. Granted 2018-12-28.

    Although the primary endpoint was not met, the effectiveness analyses described above demonstrated that the probable benefit of the device outweighs the probable risk, as discussed in the Benefit-Risk Determination section below.

Details

AI system

Workflow

The patient watched a video move around a screen five times while a camera tracked both eyes at 500 frames per second. A reference standard based on concussion symptom and cognitive scores defined each case.

Follow-up

Conditions for replication

The patient is 5 to 67 years old and within one week of a head injury. Clinicians use the result with a standard neurological assessment and do not rely on it alone. The patient has no disorder of eye movement and is not intoxicated.

Entry details

Sector

Healthcare.

Purpose

Oculogica built EyeBOX to measure and analyze eye movements as an aid in diagnosing concussion within one week of a head injury.

Not yet checked: model, release, provider, role, configuration, inputs and outputs, autonomy, human oversight, AI contribution, beneficiaries, reach, distribution and access, duration, every measurement question, every money question, ownership and control, dependencies, cross-border connections, geopolitical implications, replication attempts, availability, institutional decision, grounds, action and result, success criterion, every evidence question, observed or intended, alternative assessments, related entries, other identifiers, category mappings.

Cite this record

Peter H. Vartanian. (2026). EyeBOX missed its accuracy goals but helped rule out concussion (AIBD-0784). Artificial Intelligence Benefits Database, edition of 31 October 2026. https://petervartanian.xyz/aibd/records/AIBD-0784.html (archived at https://doi.org/10.7910/DVN/AADSII, version 4.0)

BibTeX Data Field definitions