Skip to content

Oxehealth Vital Signs read pulse and breathing from a room camera

In a study of 60 adults that FDA reviewed, Oxehealth Vital Signs measured pulse from video with an RMSD of 1.81 bpm. Breathing rate had an RMSD of 1.17 breaths/minute.

  • Not rated an organization

    The study measured accuracy against a contact device and did not count room entries or harms avoided.

Depth runs from 1, a convenience or time saved, to 5, a life or a livelihood preserved.

Status
Intended benefit.
Evidence

found, the grade of this recordstudiedreportedclaimed

An independent regulatory document supports the benefit.

Compared with

Study objectives for the root mean square difference from a contact device. Measured.

The system’s part
Not isolated. The study compared the software with a contact device. It did not compare the safety of staff or patients with and without the software. FDA reviewers describe a software algorithm that tracks pixel intensity and chest motion, and they do not say that it learned from data.
Limits
Not yet checked.
Costs and harms
Not yet checked.
Later
Not yet checked.
Benefit type
Reduced Risk
System
Oxehealth Vital Signs
Made by
Oxehealth Limited
Used by
Oxehealth (clinical accuracy study in 60 adults against a contact device)
Stage
In a pilot
Date
, when the source was published
Kind
Aggregate result
Broad benefit
Estimating the Quantifiable Characteristics of Products, Events, or Information, Monitoring Processes, Materials, or Surroundings
Collections
Scrutiny and findings
Last revised
31 October 2026

Sources

  1. Regulatory document, by authors with no tie to the system.

    U.S. Food and Drug Administration, Center for Devices and Radiological Health. De Novo Classification Request for Oxehealth Vital Signs (DEN200019): decision summary. Granted 2021-03-26.

    The pulse rate RMSD was found to be 1.81 bpm (CI 0 - 2.19 bpm). The estimated breathing rate (chest wall movements) RMSD was found to be 1.17 breaths/minute (CI 0 - 1.33) breaths/minute.

Details

AI system

Workflow

Fixed near-infrared cameras recorded each single-occupancy room. The subject was still, with enough exposed skin and chest movement visible to the camera.

Follow-up

Conditions for replication

Trained staff with a duty of care take the readings as spot checks alongside periodic checks in person. The subject is an adult who does not need intensive care or continuous monitoring of vital signs.

Entry details

Sector

Healthcare.

Purpose

Oxehealth built Oxehealth Vital Signs to let trained staff check pulse and breathing rates from video in single rooms, without entering or disturbing the patient.

Not yet checked: model, release, provider, role, configuration, inputs and outputs, autonomy, human oversight, AI contribution, beneficiaries, reach, distribution and access, duration, every measurement question, every money question, ownership and control, dependencies, cross-border connections, geopolitical implications, replication attempts, availability, institutional decision, grounds, action and result, success criterion, every evidence question, observed or intended, alternative assessments, related entries, other identifiers, category mappings.

Cite this record

Peter H. Vartanian. (2026). Oxehealth Vital Signs read pulse and breathing from a room camera (AIBD-0781). Artificial Intelligence Benefits Database, edition of 31 October 2026. https://petervartanian.xyz/aibd/records/AIBD-0781.html (archived at https://doi.org/10.7910/DVN/AADSII, version 4.0)

BibTeX Data Field definitions