AutoEF estimated how well the heart pumps within its error goal

In FDA's review of Caption Interpretation AutoEF, its estimates of ejection fraction on 186 patient studies differed from expert readings by an RMSD of 7.21. The goal was below 9.2%.

  • Not rated an organization

    The estimates were checked on stored studies, and no clinician used them to guide care.

Depth runs from 1, a convenience or time saved, to 5, a life or a livelihood preserved.

Status
Intended benefit.
Evidence

found, the grade of this recordstudiedreportedclaimed

An independent regulatory document supports the benefit.

Compared with

Pre-specified acceptance criterion for the root mean square deviation from expert measurements. Measured.

Without the system
Goal of RMSD below 9.2%
With it
RMSD 7.21 for the best available view
The system’s part
Not isolated. FDA reviewers compared the software's estimates with expert measurements of the same studies. They did not compare clinicians' reports or patient care with and without the software.
Limits
Not yet checked.
Costs and harms
Not yet checked.
Later
Not yet checked.
Benefit type
Improved Task Accuracy
System
Caption Interpretation Automated Ejection Fraction Software (AutoEF)
Made by
Caption Health, Inc.
Used by
Caption Health (retrospective validation on 186 patient studies from three US sites)
Where
Minneapolis Heart Institute, Duke University and Northwestern University, United States
Stage
Tested in a laboratory
Date
, when the source was published
Kind
Aggregate result
Broad benefit
Estimating the Quantifiable Characteristics of Products, Events, or Information, Identifying Objects, Actions, and Events
Collections
Scrutiny and findings
Last revised
31 October 2026

Sources

  1. Regulatory document, by authors with no tie to the system.

    U.S. Food and Drug Administration, Center for Devices and Radiological Health. De Novo Classification Request for Caption Interpretation Automated Ejection Fraction Software (DEN220063): decision summary. Granted 2023-02-24.

    The primary endpoint for the subject device was met with RMSD EF%[95% CIs]: 7.21[6.62, 7.74] for the best available view.

Details

AI system

Workflow

A convolutional neural network classified the clips and selected the best available views. The model that estimates ejection fraction was trained on apical and parasternal long-axis views and locked before validation. Expert cardiologists measured the reference values with the biplane Modified Simpson's method.

Follow-up

Conditions for replication

The patient is an adult with a transthoracic echocardiogram. A qualified professional checks that the images are adequate for the measurement. The clinician reviews the estimate and accepts, edits or rejects it.

Entry details

Sector

Healthcare.

Purpose

Caption Health built AutoEF to estimate left ventricular ejection fraction automatically from transthoracic ultrasound images and assist clinicians in a cardiac evaluation.

Not yet checked: model, release, provider, role, configuration, inputs and outputs, autonomy, human oversight, AI contribution, beneficiaries, reach, distribution and access, duration, every measurement question, every money question, ownership and control, dependencies, cross-border connections, geopolitical implications, replication attempts, availability, institutional decision, grounds, action and result, success criterion, every evidence question, observed or intended, alternative assessments, related entries, other identifiers, category mappings.

Cite this record

Peter H. Vartanian. (2026). AutoEF estimated how well the heart pumps within its error goal (AIBD-0778). Artificial Intelligence Benefits Database, edition of 31 October 2026. https://petervartanian.xyz/aibd/records/AIBD-0778.html (archived at https://doi.org/10.7910/DVN/AADSII, version 4.0)