Fibresolve flagged more lung fibrosis than experts in FDA review

In FDA's review, Fibresolve flagged idiopathic pulmonary fibrosis on chest CT with 55% sensitivity, against 20% for an expert panel. Its specificity was 82%, against 95%.

  • Not rated an organization

    The comparison used archived cases, and no patient was referred because of the software.

Depth runs from 1, a convenience or time saved, to 5, a life or a livelihood preserved.

Status
Intended benefit.
Evidence

found, the grade of this recordstudiedreportedclaimed

An independent regulatory document supports the benefit.

Compared with

An expert panel of five pulmonologists and thoracic radiologists reviewing the same cases. Measured.

Without the system
Panel sensitivity 20%, specificity 95%
With it
Device sensitivity 55%, specificity 82%
The system’s part
The device and an expert panel judged the same cases against the same reference diagnoses, so the difference is attributed to the software. FDA notes that the panel used stricter criteria than current practice.
Limits
Not yet checked.
Costs and harms
Not yet checked.
Later
Not yet checked.
Benefit type
Improved Task Accuracy
System
Fibresolve
Made by
Imvaria, Inc.
Used by
Imvaria (retrospective study on 300 CT cases from the Lung Tissue Research Consortium)
Where
Lung Tissue Research Consortium sites (not named in the decision summary)
Stage
Tested in a laboratory
Date
, when the source was published
Kind
Aggregate result
Broad benefit
Identifying Objects, Actions, and Events
Collections
Scrutiny and findings
Last revised
31 October 2026

Sources

  1. Regulatory document, by authors with no tie to the system.

    U.S. Food and Drug Administration, Center for Devices and Radiological Health. De Novo classification request for Fibresolve (DEN220040): decision summary. Granted 2024-01-12.

    The device appears to identify true positive IPF patients, at the expense of a high rate of false positives (specificity lower bound of 73%). This tradeoff offers a positive benefit-risk determination, considering the benefits provided by an early IPF diagnosis.

Details

AI system

Workflow

The model analyzed a single stack of axial CT slices 3 mm thick or less. Results went to clinicians as a report without images.

Follow-up

Conditions for replication

The result is used within a referral to multidisciplinary discussion or as part of one. The result is never used to rule out idiopathic pulmonary fibrosis.

Entry details

Sector

Healthcare.

Purpose

Imvaria built Fibresolve to classify lung CT scans in suspected interstitial lung disease as suggestive of idiopathic pulmonary fibrosis, as an adjunct before invasive testing.

Not yet checked: model, release, provider, role, configuration, inputs and outputs, autonomy, human oversight, beneficiaries, reach, distribution and access, duration, every measurement question, every money question, ownership and control, dependencies, cross-border connections, geopolitical implications, replication attempts, availability, institutional decision, grounds, action and result, success criterion, every evidence question, observed or intended, alternative assessments, related entries, other identifiers, category mappings.

Cite this record

Peter H. Vartanian. (2026). Fibresolve flagged more lung fibrosis than experts in FDA review (AIBD-0773). Artificial Intelligence Benefits Database, edition of 31 October 2026. https://petervartanian.xyz/aibd/records/AIBD-0773.html (archived at https://doi.org/10.7910/DVN/AADSII, version 4.0)