AI system
- System
-
EasyScan GO digital malaria microscopy device with convolutional-neural-network image analysis of thick blood films
- Model
-
A convolutional neural network that analyzes thick blood films.
- Release
-
Version 1.15.17b.
- Application developer
-
Intellectual Ventures' Global Good Fund (software and original hardware); Motic (production prototype device, software licensed to Motic)
- Role
-
The network analyzed microscope images for malaria parasites.
- Configuration
-
The deployment used software version 1.15.17b.
- Inputs and outputs
-
Not assessed
- Autonomy
-
Not assessed
- Workflow
-
EasyScan GO devices donated and set up with manufacturer training at each site. Giemsa-stained thick blood films read in parallel by two expert microscopists as reference. STARD-reported multi-center observational design across 11 countries (ClinicalTrials.gov NCT03512678).
- Human oversight
-
Not assessed
- Relation to AI
-
The machine-learning device produced the readings compared against expert microscopy, so the measured accuracy is the AI's own performance. The evidence is a single peer-reviewed study funded by the maker (Global Good Fund) with maker-employed co-authors who provided inputs to study design and training. The authors state the study did not evaluate performance under routine field operation and that all parameters were significantly affected by slide quality. No independent or outcome-level evidence was found.
- AI contribution
-
The machine-learning device produced the readings compared against expert microscopy, so the measured accuracy is the AI's own performance. The evidence is a single peer-reviewed study funded by the maker (Global Good Fund) with maker-employed co-authors who provided inputs to study design and training. The authors state the study did not evaluate performance under routine field operation and that all parameters were significantly affected by slide quality. No independent or outcome-level evidence was found.
Benefits
- Outcome identifiers
-
Not assessed
- Benefit
-
Diagnostic sensitivity was 91.1% (95% CI 88.9–92.7) and specificity 75.6% (73.1–78.0)
- Category
-
Improved Task Accuracy
- Status
-
Not assessed
- Beneficiary groups
-
People. Organizations. Society.
- Beneficiaries
-
Not assessed
- Reach
-
Not assessed
- Distribution and access
-
Not assessed
- Duration
-
Not assessed
- Expected?
-
Yes
- Costs and harms
-
Not assessed
Measurement
- Studies
-
Not assessed
- Study design
-
Not assessed
- Measurement
-
Not assessed
- Setting and sample
-
Not assessed
- Measures
-
Diagnostic sensitivity was 91.1% (95% CI 88.9–92.7) and specificity 75.6% (73.1–78.0)
- Comparison
-
expert microscopy
- Results
-
Result: diagnostic sensitivity of the EasyScan Go device was 91.1% (95% CI 88.9–92.7) and specificity was 75.6% (95% CI 73.1–78.0).
- Additional value from AI
-
Not assessed
- Limitations
-
Not assessed
- Denominators
-
Not assessed
- Materials for replication
-
Not assessed
Money
- Kind of value
-
Not assessed
- Amount
-
Not assessed
- Currency and prices
-
Not assessed
- Period and unit
-
Not assessed
- Realization
-
Not assessed
- Recipients and payers
-
Not assessed
- Gross or net
-
Not assessed
- Included costs
-
Not assessed
- Comparison
-
Not assessed
- Method
-
Not assessed
- Return and payback
-
Not assessed
People and places
- Participants
-
Developer: Intellectual Ventures' Global Good Fund (software and original hardware); Motic (production prototype device, software licensed to Motic). Operator: Research sites in 11 countries under a multi-center study with University of Oxford / Mahidol-Oxford Tropical Medicine Research Unit-affiliated investigators (research use, not routine service).
- Jurisdictions
-
Benefit: 11 sites in
Burkina Faso,
Kenya, Republic of the Congo, Senegal, South Africa, Uganda, Bangladesh,
Cambodia, Nepal,
Thailand and
Brazil. Developer:
United States. - Ownership and control
-
Not assessed
- Dependencies
-
Not assessed
- Cross-border connections
-
Not assessed
- Geopolitical implications
-
Not assessed
Follow-up
- Conditions for replication
-
Good slide quality, which the authors found significantly affected all performance parameters. Parasitemia above about 200 parasites/uL, below which sensitivity fell to 57%. Thick-film-only software (the tested version did not separate non-falciparum species).
- Replication attempts
-
2 follow-up entries. Benefit recurred: 2.
- Later findings
-
No later evaluation, approval or deployment is reported in the article.
- Availability
-
Not assessed
- Institutional decision
-
Not assessed
- Grounds
-
Not assessed
- Action and result
-
Not assessed
- Success criterion
-
Not assessed
Evidence
- Source numbers
-
AIBD-0352-source-0001–0004
- Citations
-
Numbered references below.
- Language
-
Not assessed
- Source relationships
-
Developer
- Original evidence
-
Not assessed
- Access and versions
-
Not assessed
- Supported findings
-
Not assessed
- Evidence links
-
Not assessed
- Evidence strength
-
Not assessed
- Missing evidence
-
Not assessed
Grades
- Assessed benefits
-
People. Organizations. Society.
- Depth
-
Not assessed
- Reasons
-
People: No patient-management or clinical outcome was measured. Organizations: Laboratory-level accuracy was measured under study conditions, not routine field operation. Society: No deployment at program scale is documented.
- Observed or intended
-
Not assessed
- Rubric
-
Not assessed
- Assessor and date
-
Not assessed
- Alternative assessments
-
Not assessed
- Fill
-
The makers' claim
Entry details
- Kind
-
Not assessed
- Related entries
-
Not assessed
- Field
-
Malaria diagnostics / digital microscopy. Global-health diagnostic research.
- Purpose
-
Automated detection, species identification and quantification of malaria parasites in blood films to replace or support expert microscopy. The evaluated benefit (diagnostic accuracy) is that purpose.
- Dates
-
Date basis unassessed: 2022 (precision unassessed). Updated: 21 October 2026.
- Stage
-
Not assessed
- Version
-
SHA-256: 165d995b2ec7…
References
- Other identifiers
-
Not assessed
- Category mappings
-
Not assessed
- Brief
-
Not assessed
- Revisions
-
5 revisions. Latest: 21 October 2026.
Sources
-
Peer-reviewed study Developer
Das, D., Vongpromek, R., Assawariyathipat, T., et al. (2022). Field evaluation of the diagnostic performance of EasyScan GO: a digital malaria microscopy device based on machine-learning. Malaria Journal 21, 122. doi:10.1186/s12936-022-04146-1.
The diagnostic sensitivity of EasyScan Go was 91.1% (95% CI 88.9-92.7), and specificity 75.6% (95% CI 73.1-78.0).
-
Peer-reviewed study Developer
Das, D., et al. (2022). Same article, full text (methods and limitations). Europe PMC PMC9004086.
In addition, the study did not evaluate the performance of the EasyScan Go under routine field operation.
MPH, SKM, CM, SB, CBD and CB were employees of the Global Good Fund, Intellectual Ventures at the time of study implementation
-
Peer-reviewed study Developer
Collections: Cross-border cases.