EasyScan GO reaches 91.1% sensitivity for malaria at 11 sites

Das et al. 2022: 2,250 blood films at 11 sites in 11 countries (2018–2019); EasyScan GO sensitivity 91.1%, specificity 75.6% vs expert microscopy; no patient outcome; Global Good funded.

AI system

System

EasyScan GO digital malaria microscopy device with convolutional-neural-network image analysis of thick blood films

Model

A convolutional neural network that analyzes thick blood films.

Release

Version 1.15.17b.

Application developer

Intellectual Ventures' Global Good Fund (software and original hardware); Motic (production prototype device, software licensed to Motic)

Role

The network analyzed microscope images for malaria parasites.

Configuration

The deployment used software version 1.15.17b.

Inputs and outputs

Not assessed

Autonomy

Not assessed

Workflow

EasyScan GO devices donated and set up with manufacturer training at each site. Giemsa-stained thick blood films read in parallel by two expert microscopists as reference. STARD-reported multi-center observational design across 11 countries (ClinicalTrials.gov NCT03512678).

Human oversight

Not assessed

Relation to AI

The machine-learning device produced the readings compared against expert microscopy, so the measured accuracy is the AI's own performance. The evidence is a single peer-reviewed study funded by the maker (Global Good Fund) with maker-employed co-authors who provided inputs to study design and training. The authors state the study did not evaluate performance under routine field operation and that all parameters were significantly affected by slide quality. No independent or outcome-level evidence was found.

AI contribution

The machine-learning device produced the readings compared against expert microscopy, so the measured accuracy is the AI's own performance. The evidence is a single peer-reviewed study funded by the maker (Global Good Fund) with maker-employed co-authors who provided inputs to study design and training. The authors state the study did not evaluate performance under routine field operation and that all parameters were significantly affected by slide quality. No independent or outcome-level evidence was found.

Benefits

Outcome identifiers

Not assessed

Benefit

Diagnostic sensitivity was 91.1% (95% CI 88.9–92.7) and specificity 75.6% (73.1–78.0)

Category

Improved Task Accuracy

Status

Not assessed

Beneficiary groups

People. Organizations. Society.

Beneficiaries

Not assessed

Reach

Not assessed

Distribution and access

Not assessed

Duration

Not assessed

Expected?

Yes

Costs and harms

Not assessed

Measurement

Studies

Not assessed

Study design

Not assessed

Measurement

Not assessed

Setting and sample

Not assessed

Measures

Diagnostic sensitivity was 91.1% (95% CI 88.9–92.7) and specificity 75.6% (73.1–78.0)

Comparison

expert microscopy

Results

Result: diagnostic sensitivity of the EasyScan Go device was 91.1% (95% CI 88.9–92.7) and specificity was 75.6% (95% CI 73.1–78.0).

Additional value from AI

Not assessed

Limitations

Not assessed

Denominators

Not assessed

Materials for replication

Not assessed

Money

Kind of value

Not assessed

Amount

Not assessed

Currency and prices

Not assessed

Period and unit

Not assessed

Realization

Not assessed

Recipients and payers

Not assessed

Gross or net

Not assessed

Included costs

Not assessed

Comparison

Not assessed

Method

Not assessed

Return and payback

Not assessed

People and places

Participants

Developer: Intellectual Ventures' Global Good Fund (software and original hardware); Motic (production prototype device, software licensed to Motic). Operator: Research sites in 11 countries under a multi-center study with University of Oxford / Mahidol-Oxford Tropical Medicine Research Unit-affiliated investigators (research use, not routine service).

Jurisdictions

Benefit: 11 sites in Burkina Faso, Kenya, Republic of the Congo, Senegal, South Africa, Uganda, Bangladesh, Cambodia, Nepal, Thailand and Brazil. Developer: United States.

Ownership and control

Not assessed

Dependencies

Not assessed

Cross-border connections

Not assessed

Geopolitical implications

Not assessed

Follow-up

Conditions for replication

Good slide quality, which the authors found significantly affected all performance parameters. Parasitemia above about 200 parasites/uL, below which sensitivity fell to 57%. Thick-film-only software (the tested version did not separate non-falciparum species).

Replication attempts

2 follow-up entries. Benefit recurred: 2.

Later findings

No later evaluation, approval or deployment is reported in the article.

Availability

Not assessed

Institutional decision

Not assessed

Grounds

Not assessed

Action and result

Not assessed

Success criterion

Not assessed

Evidence

Source numbers

AIBD-0352-source-0001–0004

Citations

Numbered references below.

Language

Not assessed

Source relationships

Developer

Original evidence

Not assessed

Access and versions

Not assessed

Supported findings

Not assessed

Evidence strength

Not assessed

Missing evidence

Not assessed

Grades

Assessed benefits

People. Organizations. Society.

Depth

Not assessed

Reasons

People: No patient-management or clinical outcome was measured. Organizations: Laboratory-level accuracy was measured under study conditions, not routine field operation. Society: No deployment at program scale is documented.

Observed or intended

Not assessed

Rubric

Not assessed

Assessor and date

Not assessed

Alternative assessments

Not assessed

Fill

The makers' claim

Entry details

Kind

Not assessed

Field

Malaria diagnostics / digital microscopy. Global-health diagnostic research.

Purpose

Automated detection, species identification and quantification of malaria parasites in blood films to replace or support expert microscopy. The evaluated benefit (diagnostic accuracy) is that purpose.

Dates

Date basis unassessed: 2022 (precision unassessed). Updated: 21 October 2026.

Stage

Not assessed

Version

SHA-256: 165d995b2ec7…

References

Other identifiers

Not assessed

Category mappings

Not assessed

Brief

Not assessed

Revisions

5 revisions. Latest: 21 October 2026.

Sources

  1. Peer-reviewed study Developer

    Das, D., Vongpromek, R., Assawariyathipat, T., et al. (2022). Field evaluation of the diagnostic performance of EasyScan GO: a digital malaria microscopy device based on machine-learning. Malaria Journal 21, 122. doi:10.1186/s12936-022-04146-1.

  2. Peer-reviewed study Developer

    Das, D., et al. (2022). Same article, full text (methods and limitations). Europe PMC PMC9004086.

  3. Peer-reviewed study Developer

    Das, D., Vongpromek, R., Assawariyathipat, T., et al. (2022). Field evaluation of the diagnostic performance of EasyScan GO: a digital malaria microscopy device based on machine-learning. Malaria Journal 21, 122. doi:10.1186/s12936-022-04146-1. Full text as served by the Europe PMC REST API (PMC9004086).