An AI eye exam increased screening among young people with diabetes

In the ACCESS trial, all 81 young people offered IDx-DR completed an eye exam, compared with 18 of 82 controls referred to an eye-care provider.

AI system

System

IDx-DR, now sold as LumineticsCore. [1]

Model

The locked diagnostic algorithms within IDx-DR. The trial does not name a separate model family or release. [1]

Release

The paper and supplement do not identify the installed software release. [1]

Provider

Digital Diagnostics, United States. [1]

Role

The software checked image quality and assessed photographs for diabetic eye disease. [1]

Configuration

The device used two retinal views and allowed up to three attempts to obtain suitable images. The trial does not name the camera model. [1]

Inputs and outputs

Two color retinal photographs centered on the fovea and optic disc were assessed. The output indicated disease, no disease, or insufficient image quality. [1]

Autonomy

The software produced its result without a clinician interpreting the photographs first. The research team nevertheless obtained specialist overreads for safety. [1]

Workflow

Clinic staff photographed the retina during the diabetes visit. Participants received the result immediately and were referred for an eye-care visit when indicated. [1]

Human oversight

A board-certified retina specialist reviewed every AI image for safety. An independent monitoring board oversaw the study. [1]

Relation to AI

IDx-DR assessed retinal photographs and returned a screening result during the diabetes visit. The trial compared this on-site service with referral elsewhere. A retina specialist also reviewed the images for safety. A coauthor had financial interests in the manufacturer, although he did not have access to the underlying data.

AI contribution

IDx-DR made an eye exam available at the diabetes visit. The trial tests this entire on-site service against referral elsewhere. [1]

Benefits

Outcome identifiers

AIBD-0067-outcome-01 [1]

Benefit

More participants completed screening when an AI eye exam was offered during their diabetes visit. [1]

Category

Improved Physical Health

Status

Observed in a randomized trial. [1]

Beneficiary groups

People.

Beneficiaries

Young people with type 1 or type 2 diabetes who were due for a diabetic eye exam. [1]

Reach

All 81 AI participants and 18 of 82 controls completed screening. The 25 positive AI screens were not all confirmed diagnoses. [1]

Distribution and access

The trial included a racially diverse population, and 47% of participants had Medicaid coverage. It was not large enough to establish equal effects across every subgroup. [1]

Duration

Screening and referral attendance were followed for six months. [1]

Expected?

Yes

Costs and harms

The AI produced positive screening results that required further assessment. Its reported specificity was 78.9% against specialist review. A positive screen was not a confirmed diagnosis. [1]

Measurement

Studies

NCT05131451 [1]

Study design

An unmasked randomized trial, with allocation concealed until assignment. Randomization used blocks of four or six and was stratified by clinic. [1]

Measurement

The primary outcome was documented completion of a diabetic eye exam within six months. Researchers also tracked referral attendance and participant satisfaction. [1]

Setting and sample

The trial randomized 164 participants at two Baltimore clinics. After one control was excluded, the analysis included 81 AI participants and 82 controls. [1]

Measures

Completion of a diabetic eye exam within six months, with referral attendance as a secondary outcome. [1]

Comparison

Scripted education and referral to an eye-care provider, with a printed handout. [1]

Results

Screening completion was 100% with AI and 22% with referral, a difference of 78 percentage points (95% CI, 69–87). [1]

Additional value from AI

Randomization supports the effect of offering an immediate AI exam rather than a referral. It does not separate the algorithm from convenience and access at the clinic. [1]

Limitations

The trial measured access to screening, not later vision. The referral comparison used 25 AI-positive participants versus the entire control group, so those secondary denominators are not equivalent. [1]

Denominators

Participants enrolled from 24 November 2021 to 6 June 2022. Follow-up ended on 6 December 2022. An AI exam generally took five to ten minutes. [1]

Materials for replication

The protocol and supplement describe the procedure. De-identified data require an approved proposal and data-use agreement. The commercial diagnostic code is not published. [1]

Money

Kind of value

A gift card for study participation. This was a transfer, not value that AI created. [1]

Amount

US$25 per participant, plus a parking pass whose value was not reported. [1]

Currency and prices

U.S. dollars during the 2021–2022 study. No inflation adjustment was reported. [1]

Period and unit

Per participant for study participation. [1]

Realization

The authors report that participants received the gift card and parking pass. [1]

Recipients and payers

Study participants received compensation. The article does not identify the precise paying account. [1]

Gross or net

The US$25 is the face value of a gift card. It is not net benefit from screening. [1]

Included costs

The article does not report the full cost of the service, specialist review, staff time, or parking passes. [1]

Comparison

No comparison of financial outcomes between the randomized groups was reported. [1]

Method

The source states the gift card’s face value. It does not value health outcomes. [1]

Return and payback

No financial return or payback period was estimated. [1]

People and places

Participants

Johns Hopkins conducted the trial at Johns Hopkins Hospital and Mount Washington Pediatric Hospital. Digital Diagnostics supplied the diagnostic product. [1]

Jurisdictions

Both clinics were in Baltimore, Maryland, United States. The manufacturer was also based in the United States. [1]

Ownership and control

The reviewed clinical sources identify the organizations involved, but do not establish their complete ownership or control arrangements. [1]

Dependencies

The service required a retinal camera, trained clinic staff, IDx-DR, specialist safety review, and a referral route. [1]

Cross-border connections

The reviewed sources do not document international transfers of patient data or quantify cross-border financial flows. [1]

Geopolitical implications

Not assessed

Follow-up

Conditions for replication

The tested service combined an immediate eye exam, clinic staff, follow-up referrals, and specialist safety review for young participants. [1]

Replication attempts

The reviewed sources do not report an independent repetition of this clinical result. [1]

Later findings

The study followed eye-exam completion and referral attendance for six months. It did not report later vision outcomes.

Availability

LumineticsCore is commercially available. The trial’s pediatric use was investigational because the FDA indication covered adults aged 22 or older. [1]

Institutional decision

Johns Hopkins evaluated AI screening in young people below the device’s FDA-indicated age. [1]

Grounds

The study received IRB approval, an investigational device exemption, and independent safety monitoring. [1]

Action and result

Staff used the device in the trial and arranged retina-specialist review of every AI image. [1]

Success criterion

The prespecified primary outcome was completion of an eye exam within six months. [1]

Evidence

Source numbers

1 retained sources.

Citations

Original citations and the passages that support each finding. [1]

Language

English. No translation was used. [1]

Source relationships

Academic investigators conducted the trial. One coauthor was an investor, director, and consultant of Digital Diagnostics, but had no access to the underlying data. [1]

Original evidence

The original studies and their supporting materials are linked to the findings that they support. [1]

Access and versions

Access and version details are listed for each reviewed source. [1]

Supported findings

An on-site IDx-DR exam increased screening completion from 22% to 100%. [1]

Evidence strength

A peer-reviewed randomized trial by academic investigators, with a coauthor who held financial interests in the manufacturer. [1]

Missing evidence

The paper does not establish later vision outcomes, the device’s exact software release, or a separate causal effect of the algorithm. [1]

Grades

Assessed benefits

The benefit described here, for the beneficiary groups shown.

Depth

Depth 2 for improved access to completed screening.

Reasons

Depth 2 for improved access to completed screening.

Observed or intended

The depth describes the reported result. Evidence strength is assessed separately.

Rubric

AIBD’s five-level depth rubric, reviewed on 25 October 2026.

Assessor and date

AIBD, 25 October 2026.

Alternative assessments

A null grade would omit the measured improvement. Grades 3–5 would require a lasting capability or health benefit that this trial did not establish.

Fill

A randomized academic evaluation with a manufacturer-affiliated coauthor. The outline is retained under the current independence rule.

Entry details

Kind

A randomized trial of on-site screening versus referral. [1]

Field

Healthcare - diabetic retinopathy screening. Autonomous AI screening in pediatric diabetes care.

Purpose

IDx-DR/LumineticsCore is built to autonomously detect more-than-mild diabetic retinopathy from fundus images at the point of care. Increasing screening completion is its intended benefit.

Dates

Enrollment began on 24 November 2021, and follow-up ended on 6 December 2022. [1]

Stage

Investigational use in a pediatric diabetes trial. [1]

Version

SHA-256: f6eae30ef3c1…

References

Other identifiers

NCT05131451 [1]

Category mappings

Not assessed

Brief

Not assessed

Revisions

7 revisions. Latest: 25 October 2026.

Sources

  1. Peer-reviewed study Evaluator

    Risa M. Wolf et al., “Autonomous Artificial Intelligence Increases Screening and Follow-up for Diabetic Retinopathy in Youth: The ACCESS Randomized Control Trial,” Nature Communications 15 (2024): 421, doi:10.1038/s41467-023-44676-z. Figures 1–2, Methods, and Supplementary Note 3.

    Randomized screening results, denominators, age limits, procedures, and conflicts. Figure 1; Tables 1–4; Methods; Data and Code availability; Author contributions; Competing interests.

    Original academic evaluation with a manufacturer-affiliated coauthor.

    Supplement to the same trial. Image procedure, investigational use, and specialist safety review. Supplementary Note 3 and supplementary tables.

    Supplement to the same trial.

    Sponsor-supplied trial registration. Study registration and ethics identifier. Protocol identification, status, locations, eligibility, and outcomes.

    Sponsor-supplied trial registration.