AI system
- System
-
Stentrode, with a custom decoder and Tobii Dynavox software. [1]
- Model
-
A support-vector machine with a radial-basis kernel for participant 1, and a manually tuned threshold classifier for participant 2. [1]
- Release
-
No software release or trained-weight identifier is reported. NCT03834857 identifies the trial, not the model. [1]
- Provider
-
Synchron supplied Stentrode. The research team developed the decoding method, and Tobii Dynavox supplied eye tracking and computer-access software. [1]
- Role
-
The decoder generated clicks from attempted movements. Eye tracking moved the cursor. [1]
- Configuration
-
The first participant used a trained SVM. The second used a threshold that changed during typing. Both produced short and held clicks. [1]
- Inputs and outputs
-
Electrical signals recorded from a cerebral vein were converted into no click, a short click, or a held click. [1]
- Autonomy
-
Participants controlled selections through attempted movements. The system did not choose messages or purchases for them. [1]
- Workflow
-
A caregiver attached the receiver and opened the software. The participant then used eye movements and brain-generated clicks to operate a computer. [1]
- Human oversight
-
Clinicians selected and implanted participants, monitored safety, and managed antiplatelet treatment. Researchers tuned decoders, and caregivers helped with setup at home. [1]
- Relation to AI
-
Participant 1 used a support-vector machine trained on movement attempts. Participant 2 used a manually tuned threshold classifier. Both converted brain signals into clicks, while an eye tracker handled cursor movement. The measured benefit belonged to the combined system.
- AI contribution
-
The learned decoder supplied the click channel for one of the first two participants. The other used a threshold rule. Eye tracking and commercial software supplied the rest of the interface. [1]
Benefits
- Outcome identifiers
-
AIBD-0052-outcome-01 [1]
- Benefit
-
Participants regained the ability to select items and perform everyday computer tasks at home. [1]
- Category
-
Improved Accessibility
- Status
-
Demonstrated in a small, uncontrolled clinical study. [1]
- Beneficiary groups
-
People.
- Beneficiaries
-
People with severe upper-limb paralysis who retained suitable brain activity and venous anatomy. [1]
- Reach
-
Four unique implanted participants are described across both reports. The initial two are included in the later four. [1] [2]
- Distribution and access
-
The four implanted participants were men of European ancestry. The small, selected group does not establish broad suitability. [1]
- Duration
-
The original report covered 12 months for participant 1 and three months for participant 2. The later paper followed four implanted participants for 12 months each. [1]
- Expected?
-
Yes
- Costs and harms
-
Implantation required anesthesia and antiplatelet treatment. One participant fainted briefly after surgery. The later four-person report recorded eight mild device effects. [1]
Measurement
- Studies
-
NCT03834857 [1]
- Study design
-
A prospective, open-label feasibility study without a control group. The later report includes the first two participants. [1]
- Measurement
-
Researchers measured typing and selection accuracy, observed everyday computer tasks, recorded adverse events, and checked vessel patency and device position. [1]
- Setting and sample
-
Two implanted participants appeared in the first report. The later study enrolled five people, implanted four, and excluded one because of venous anatomy. [1]
- Measures
-
Correct characters per minute, selection accuracy, completion of everyday tasks, serious device-related events, vessel occlusion, and migration. [1]
- Comparison
-
There was no concurrent control. The comparison with the Utrecht interface involved different participants and protocols. [1]
- Results
-
The first two participants typed 13.81 and 20.10 correct characters per minute, with selection accuracy above 92%. [1] [2]
- Additional value from AI
-
The study demonstrates use of the combined implant, decoder, eye tracker, and software. It does not isolate the benefit of machine learning. [1]
- Limitations
-
The cohort was small and selected, and decoders differed between participants. Without a control group, superiority over other assistive systems cannot be established. [1]
- Denominators
-
The first participants began home use without research staff on days 86 and 71 after implantation. Caregiver assistance with setup remained necessary. [1]
- Materials for replication
-
The 2020 supplement gives participant-specific filters, training splits, and click rules. It does not publish exact trained weights or a software release. [1]
Money
- Currency and prices
-
The reviewed sources do not report a monetary valuation or financial return. [1] [2]
- Period and unit
-
The reviewed sources do not report a monetary valuation or financial return. [1] [2]
- Recipients and payers
-
The reviewed sources do not report a monetary valuation or financial return. [1] [2]
People and places
- Participants
-
Synchron and University of Melbourne researchers developed the interface. The Royal Melbourne Hospital implanted and followed participants, who used it at home. [1]
- Jurisdictions
-
The first clinical study and home use took place in Melbourne,
Australia. Synchron was based in the
United States and worked with Australian investigators. [1] - Ownership and control
-
The reviewed clinical sources identify the organizations involved, but do not establish their complete ownership or control arrangements. [1]
- Dependencies
-
The system required implantation, a chest receiver, a computer, eye tracking, individual training, caregiver support, and clinical follow-up. [1]
- Cross-border connections
-
The reviewed sources do not document international transfers of patient data or quantify cross-border financial flows. [1]
- Geopolitical implications
-
Not assessed
Follow-up
- Conditions for replication
-
Suitable anatomy, preserved motor signals, training, caregiver setup, and continuing clinical care were required. [1]
- Replication attempts
-
The later four-person report includes the original two participants. [2]
- Later findings
-
The four-person follow-up found no serious device-related events over 12 months. Eight mild device effects occurred, and one participant later demonstrated texting without eye tracking. [2]
- Availability
-
The reports concern an investigational device. They do not establish routine commercial access at the time of the study. [1]
- Institutional decision
-
Australian investigators tested the safety and feasibility of an endovascular brain–computer interface. [1] [2]
- Grounds
-
The study received hospital ethics approvals and authorization under Australia’s Clinical Trial Notification scheme. [1] [2]
Evidence
- Source numbers
-
2 retained sources.
- Language
-
English. No translation was used. [1]
- Source relationships
-
Synchron funded and participated in the later study. An independent safety monitor reviewed safety data, but this does not make the whole efficacy study independent. [2]
- Original evidence
-
The original studies and their supporting materials are linked to the findings that they support. [1] [2]
- Supported findings
-
Stentrode supported independent computer tasks after caregiver setup, with eye tracking for cursor navigation. [1]
- Evidence links
-
The first paper documents the two decoders and home tasks. The later paper adds two participants and longer safety follow-up. [1] [2]
- Evidence strength
-
Original peer-reviewed studies by the developers and clinical team. The four-person follow-up is an extension, not an independent replication. [1]
- Missing evidence
-
The studies do not establish a financial return, a broadly representative effect, or one fixed decoder used by every participant. [1]
Grades
- Assessed benefits
-
The benefit described here, for the beneficiary groups shown.
- Depth
-
Depth 3 for a restored ability that participants used at home.
- Reasons
-
Depth 3 for a restored ability that participants used at home.
- Observed or intended
-
The depth describes the reported result. Evidence strength is assessed separately.
- Rubric
-
AIBD’s five-level depth rubric, reviewed on 25 October 2026.
- Assessor and date
-
AIBD, 25 October 2026.
- Alternative assessments
-
Depth 2 understates the restored practical ability. A separate societal grade would require a demonstrated societal benefit.
- Fill
-
Peer-reviewed studies by the developers and clinical team. The outline is retained.
Entry details
- Kind
-
An early feasibility study of an implanted brain–computer interface. [1]
- Related entries
-
AIBD-0265 describes the Utrecht interface cited as a historical comparison. The studies were not head-to-head.
- Field
-
Neurotechnology / assistive communication. People with ALS and severe upper-limb paralysis.
- Purpose
-
Restore digital device control for people with severe paralysis without open-brain surgery
- Dates
-
The first implants occurred in mid-2019 and early 2020. The initial report appeared on 28 October 2020. [1] [2]
- Stage
-
Investigational implantation and supervised training, followed by use at home. [1]
- Version
-
SHA-256: 1bedfa433679…
References
- Other identifiers
-
NCT03834857 [1]
- Category mappings
-
Not assessed
- Brief
-
Not assessed
- Revisions
-
6 revisions. Latest: 25 October 2026.
Sources
-
Peer-reviewed study Developer
Oxley TJ, Yoo PE, Rind GS, et al. (2021). Motor neuroprosthesis implanted with neurointerventional surgery improves capacity for activities of daily living tasks in severe paralysis: first in-human experience. Journal of NeuroInterventional Surgery 13(2):102-108.
Original home-use outcomes, distinct decoders, training, caregiver setup, and safety. Results; Methods; Supplementary Appendix, pp. 3–10.
Original first-in-human report by developers and clinicians.
Supplement to the initial report. Detailed decoder architecture, training splits, click logic, and patient-reported outcomes. neurintsurg-2020-016862supp001.pdf, pp. 3–10.
Supplement to the initial report.
Sponsor-supplied registry entry. Original registration and study sponsor. Identification, status, design, and eligibility.
Sponsor-supplied registry entry.
Completion of IADL tasks including text messaging, online shopping and managing finances independently was demonstrated in both participants.
-
Peer-reviewed study Developer
Peter Mitchell et al., “Assessment of Safety of a Fully Implanted Endovascular Brain-Computer Interface for Severe Paralysis in 4 Patients: The Stentrode With Thought-Controlled Digital Switch (SWITCH) Study,” JAMA Neurology 80, no. 3 (2023): 270–278, doi:10.1001/jamaneurol.2022.4847. Published online January 9, 2023.
Follow-up of four implanted participants, overlap, mild adverse effects, and independent safety-monitor role. Methods; Results; Table 2; Funding and sponsor role.
Developer-funded extension of the same trial.
Supplement to the later paper. Referenced for additional methods; no uninspected details are used. 2023 Supplement 1.
Supplement to the later paper.
Collections: Cross-border cases.