{"id":"AIBD-1031","title":"Canary Islands assessment service SESCS finds that the G-STRIDE gait sensor may detect fall risk better than the Timed Up and Go test","summary":"SESCS found that, in one study, the G-STRIDE gait sensor identified more older people at risk of falling than the Timed Up and Go test and about as many as a gait-speed test.","date_observed":"2026-08-06","date_basis":"source publication","date_precision":"day","benefit_status":"mixed","source_collection":null,"source_entry":null,"account":"The Evaluation Service of the Canary Islands Health Service (SESCS) assessed wearable devices for fall risk and fall detection in people aged 65 and over. It published the assessment for Spain's national health system on 6 August 2026. Its review found 16 studies but none that measured falls, injuries or quality of life. Only one study gave the data to compute sensitivity and specificity. In that Spanish study of 157 people over 70, the G-STRIDE sensor had a sensitivity of 77% and a specificity of 83%. The Timed Up and Go test had 51% and 83%. A gait-speed test had 75% and 79%. SESCS describes G-STRIDE as a shoe-mounted inertial sensor that analyzes gait with a machine-learning algorithm. It concludes that the results suggest that G-STRIDE detects fall risk better than the Timed Up and Go test and much like the gait-speed test. The study classified people by falls in the past 12 months. SESCS rated the evidence as very low in certainty. It could not recommend the devices and suggested their use in research or pilot projects.","category":"aggregate result","unexpected":false,"purpose":"An academic consortium of Universidad Rey Juan Carlos and collaborators built G-STRIDE to analyze gait with a shoe-mounted inertial sensor and estimate the risk of falling in older people.","system":"G-STRIDE (shoe-mounted inertial sensor with a machine-learning gait algorithm)","model_version":null,"developer":"Academic consortium of Universidad Rey Juan Carlos and collaborators","deployer":"The Spanish validation study that SESCS used, with 157 people over 70","location":"Spain","domain":"Healthcare (fall prevention in older people)","segment":null,"benefit_type":"Improved Task Accuracy","work_activities":["4.A.2.a.4","4.A.1.b.3"],"taxonomy_note":null,"registry_ids":{},"ratings":[{"group":"organizational","depth":null,"reason":"The gain over the Timed Up and Go test comes from one research study that classified past fallers, with no measure of falls prevented."}],"ai_role":{"status":"assessed","statement":"The study tested G-STRIDE, the Timed Up and Go test and a gait-speed test in the same 157 people, which isolates the device's accuracy against tests in use. The device combines the sensor with a machine-learning algorithm, whose part the study does not separate."},"evidence":[{"kind":"assessment report","citation":"Servicio de Evaluación del Servicio Canario de la Salud (SESCS). Sistemas de telemonitorización portátiles para la evaluación del riesgo de caídas y la detección de caídas en personas de 65 o más años. Informes de Evaluación de Tecnologías Sanitarias, Ministerio de Sanidad, 2026. Resumen ejecutivo (Resultados, Coste-efectividad y análisis económico).","title":"Sistemas de telemonitorización portátiles para la evaluación del riesgo de caídas y la detección de caídas en personas de 65 o más años","url":"https://sescs.es/wp-content/uploads/2026/08/114_2025_SESCS_Telemonitorizacion_Caidas_DEF_NIPO.pdf","date":"2026-08-06","written_by":"independent","quote":"Los resultados sugieren que G-STRIDE mejora la detección del riesgo de caída respecto al test TUG, aunque su rendimiento es muy similar al test GS."}],"replication":{"produced_by":["An inertial measurement unit on the foot recorded gait parameters while people walked without restrictions.","A machine-learning algorithm classified fall risk from the gait parameters."],"must_hold":["A clinician must interpret the result in a consultation, as with the conventional tests.","Sensor and software costs must stay low, since the cost per extra correct case was lowest with large-scale purchase."],"echoes":[]},"paired_incident":null,"headline":"*G-STRIDE* spots more older people at risk of falls than the TUG test","metric":"sensitivity of 77%","comparator":{"against":"The Timed Up and Go test and a gait-speed test in the same people","before":"test TUG, que tuvo una sensibilidad de 51% (IC95%: 39, 62)","after":"el dispositivo portátil G-STRIDE obtuvo una sensibilidad de 77% (IC95%: 55, 85)","basis":"measured","quote":"En este estudio, el dispositivo portátil G-STRIDE obtuvo una sensibilidad de 77% (IC95%: 55, 85) y una especificidad de 83% (IC95%: 73, 91), frente al test TUG, que tuvo una sensibilidad de 51% (IC95%: 39, 62)","url":"https://sescs.es/wp-content/uploads/2026/08/114_2025_SESCS_Telemonitorizacion_Caidas_DEF_NIPO.pdf"},"later":null,"value":null,"developer_country":"Spain","deployer_country":"Spain","unrated_because":"The only comparison comes from one research study of past falls, with no study of falls, injuries or care with the device.","brief_notes":null,"brief":null,"revisions":[{"date":"2026-11-01","change":"Added the record from the finding of the Evaluation Service of the Canary Islands Health Service (SESCS).","source":"https://sescs.es/wp-content/uploads/2026/08/114_2025_SESCS_Telemonitorizacion_Caidas_DEF_NIPO.pdf","editor":"AIBD"}],"confidence":"found","shade":"high","claim_status":"Found by an assessment body","claim_basis":"Sources: one independent assessment report. The cited evidence includes no independent study or report.","case_jurisdictions":["Spain"],"profile":{"case_id":"AIBD-1031","catalog_version":"1.0.0","source_fingerprint":"1b65e597c6e25861d310da5b6d257c5522f5c4e614b9edc09c24e128aa09a7e6","collection_refs":[{"collection_id":"aibd","version":"2026-10-31"},{"collection_id":"querist","version":"2026-10-25"}],"source_ids":["AIBD-1031-source-0001"],"families":{"strategic assessments":{"state":"unresolved","provenance":{"kind":"curatorial","basis":"Editorial assignment.","record_fields":[],"source_refs":[]},"reason":"Editorial assignment.","review_state":"not_assessed","coverage":"partial","record_ids":[]},"evaluations":{"state":"unresolved","provenance":{"kind":"curatorial","basis":"Editorial assignment.","record_fields":[],"source_refs":[]},"reason":"Editorial assignment.","review_state":"not_assessed","coverage":"partial","record_ids":[]},"monetary observations":{"state":"unresolved","provenance":{"kind":"curatorial","basis":"Editorial assignment.","record_fields":[],"source_refs":[]},"reason":"Editorial assignment.","review_state":"not_assessed","coverage":"partial","record_ids":[]},"institutional decisions":{"state":"unresolved","provenance":{"kind":"curatorial","basis":"Editorial assignment.","record_fields":[],"source_refs":[]},"reason":"Editorial assignment.","review_state":"not_assessed","coverage":"partial","record_ids":[]}},"fields":{"actors[]":{"state":"value_present","value":{"kind":"record_fields","fields":{"developer":"Academic consortium of Universidad Rey Juan Carlos and collaborators","deployer":"The Spanish validation study that SESCS used, with 157 people over 70"}},"provenance":{"kind":"legacy_record","basis":"Editorial assignment.","record_fields":["developer","deployer"],"source_refs":[]},"review_state":"record_mapped","coverage":"partial"},"actors[].ownership_control":{"state":"unresolved","provenance":{"kind":"curatorial","basis":"Editorial assignment.","record_fields":[],"source_refs":[]},"reason":"This is an explicit unresolved assignment: the sources on record do not settle this question.","review_state":"not_assessed","coverage":"partial"},"ai_contribution":{"state":"value_present","value":{"kind":"record_fields","fields":{"ai_role":{"status":"assessed","statement":"The study tested G-STRIDE, the Timed Up and Go test and a gait-speed test in the same 157 people, which isolates the device's accuracy against tests in use. The device combines the sensor with a machine-learning algorithm, whose part the study does not separate."}}},"provenance":{"kind":"legacy_record","basis":"Editorial assignment.","record_fields":["ai_role"],"source_refs":[]},"review_state":"record_mapped","coverage":"partial"},"ai_relationship":{"state":"value_present","value":{"kind":"record_fields","fields":{"category":"aggregate result","ai_role":{"status":"assessed","statement":"The study tested G-STRIDE, the Timed Up and Go test and a gait-speed test in the same 157 people, which isolates the device's accuracy against tests in use. The device combines the sensor with a machine-learning algorithm, whose part the study does not separate."}}},"provenance":{"kind":"legacy_record","basis":"Editorial assignment.","record_fields":["ai_role","category"],"source_refs":[]},"review_state":"record_mapped","coverage":"partial"},"assessments[].alternatives":{"state":"unresolved","provenance":{"kind":"curatorial","basis":"Editorial assignment.","record_fields":[],"source_refs":[]},"reason":"This is an explicit unresolved assignment: the sources on record do not settle this question.","review_state":"not_assessed","coverage":"partial"},"assessments[].assessor_date":{"state":"unresolved","provenance":{"kind":"curatorial","basis":"Editorial assignment.","record_fields":[],"source_refs":[]},"reason":"This is an explicit unresolved assignment: the sources on record do not settle this question.","review_state":"not_assessed","coverage":"partial"},"assessments[].depth":{"state":"unresolved","provenance":{"kind":"curatorial","basis":"Editorial assignment.","record_fields":[],"source_refs":[]},"reason":"This is an explicit unresolved assignment: the sources on record do not settle this question.","review_state":"not_assessed","coverage":"partial"},"assessments[].display_confidence":{"state":"value_present","value":{"kind":"structured","data":{"confidence":"high","evidence_word":"found","claim_status":"Found by an assessment body","basis":"Sources: one independent assessment report. The cited evidence includes no independent study or report.","interpretation":"Existing display rule preserved; the richer outcome-specific confidence question still needs review."}},"provenance":{"kind":"curatorial","basis":"Editorial assignment.","record_fields":[],"source_refs":[]},"review_state":"record_mapped","coverage":"partial"},"assessments[].rationale":{"state":"value_present","value":{"kind":"record_fields","fields":{"ratings":[{"group":"organizational","reason":"The gain over the Timed Up and Go test comes from one research study that classified past fallers, with no measure of falls prevented."}]}},"provenance":{"kind":"legacy_record","basis":"Editorial assignment.","record_fields":["ratings"],"source_refs":[]},"review_state":"record_mapped","coverage":"partial"},"assessments[].rubric_version":{"state":"unresolved","provenance":{"kind":"curatorial","basis":"Editorial assignment.","record_fields":[],"source_refs":[]},"reason":"This is an explicit unresolved assignment: the sources on record do not settle this question.","review_state":"not_assessed","coverage":"partial"},"assessments[].status_basis":{"state":"unresolved","provenance":{"kind":"curatorial","basis":"Editorial assignment.","record_fields":[],"source_refs":[]},"reason":"This is an explicit unresolved assignment: the sources on record do not settle this question.","review_state":"not_assessed","coverage":"partial"},"assessments[].target":{"state":"value_present","value":{"kind":"record_fields","fields":{"ratings":[{"group":"organizational","reason":"The gain over the Timed Up and Go test comes from one research study that classified past fallers, with no measure of falls prevented."}]}},"provenance":{"kind":"legacy_record","basis":"Editorial assignment.","record_fields":["ratings"],"source_refs":[]},"review_state":"record_mapped","coverage":"partial"},"availability":{"state":"unresolved","provenance":{"kind":"curatorial","basis":"Editorial assignment.","record_fields":[],"source_refs":[]},"reason":"This is an explicit unresolved assignment: the sources on record do not settle this question.","review_state":"not_assessed","coverage":"partial"},"brief":{"state":"unresolved","provenance":{"kind":"curatorial","basis":"Editorial assignment.","record_fields":[],"source_refs":[]},"reason":"This is an explicit unresolved assignment: the sources on record do not settle this question.","review_state":"not_assessed","coverage":"partial"},"case_kind":{"state":"unresolved","provenance":{"kind":"curatorial","basis":"Editorial assignment.","record_fields":[],"source_refs":[]},"reason":"This is an explicit unresolved assignment: the sources on record do not settle this question.","review_state":"not_assessed","coverage":"partial"},"case_links[]":{"state":"unresolved","provenance":{"kind":"curatorial","basis":"Editorial assignment.","record_fields":[],"source_refs":[]},"reason":"This is an explicit unresolved assignment: the sources on record do not settle this question.","review_state":"not_assessed","coverage":"partial"},"claims[].evidence_links":{"state":"unresolved","provenance":{"kind":"curatorial","basis":"Editorial assignment.","record_fields":[],"source_refs":[]},"reason":"This is an explicit unresolved assignment: the sources on record do not settle this question.","review_state":"not_assessed","coverage":"partial"},"claims[].evidence_strength":{"state":"value_present","value":{"kind":"structured","data":{"answer":"Found by an assessment body. Sources: one independent assessment report. The cited evidence includes no independent study or report.","evidence_word":"found","claim_status":"Found by an assessment body"}},"provenance":{"kind":"source_review","basis":"Editorial assignment.","record_fields":["evidence"],"source_refs":["AIBD-1031-source-0001"],"reviewed_on":"2026-10-29"},"review_state":"reviewed","coverage":"complete","display":{"answer":"Found by an assessment body. Sources: one independent assessment report. The cited evidence includes no independent study or report.","inferred":false}},"claims[].missingness":{"state":"unresolved","provenance":{"kind":"curatorial","basis":"Editorial assignment.","record_fields":[],"source_refs":[]},"reason":"This is an explicit unresolved assignment: the sources on record do not settle this question.","review_state":"not_assessed","coverage":"partial"},"claims[].proposition":{"state":"unresolved","provenance":{"kind":"curatorial","basis":"Editorial assignment.","record_fields":[],"source_refs":[]},"reason":"This is an explicit unresolved assignment: the sources on record do not settle this question.","review_state":"not_assessed","coverage":"partial"},"cross_border_flows[]":{"state":"unresolved","provenance":{"kind":"curatorial","basis":"Editorial assignment.","record_fields":[],"source_refs":[]},"reason":"This is an explicit unresolved assignment: the sources on record do not settle this question.","review_state":"not_assessed","coverage":"partial"},"decisions[].action_outcome":{"state":"unresolved","review_state":"not_assessed","coverage":"partial","family":"institutional decisions"},"decisions[].grounds":{"state":"unresolved","review_state":"not_assessed","coverage":"partial","family":"institutional decisions"},"decisions[].institution_task":{"state":"unresolved","review_state":"not_assessed","coverage":"partial","family":"institutional decisions"},"decisions[].success_criterion":{"state":"unresolved","review_state":"not_assessed","coverage":"partial","family":"institutional decisions"},"deployment_stage":{"state":"value_present","value":{"kind":"structured","data":{"classification":"laboratory test","answer":"The system was tested in a laboratory. SESCS suggests using the device only in research or pilot projects, after one research validation study.","scope":"The period and activity described in the case."}},"provenance":{"kind":"source_review","basis":"Editorial assignment.","record_fields":["account","evidence"],"source_refs":["AIBD-1031-source-0001"],"reviewed_on":"2026-10-29"},"review_state":"reviewed","coverage":"complete","display":{"answer":"The system was tested in a laboratory. SESCS suggests using the device only in research or pilot projects, after one research validation study.","inferred":false}},"economics[].amount":{"state":"unresolved","review_state":"not_assessed","coverage":"partial","family":"monetary observations"},"economics[].costs[]":{"state":"unresolved","review_state":"not_assessed","coverage":"partial","family":"monetary observations"},"economics[].counterfactual":{"state":"unresolved","review_state":"not_assessed","coverage":"partial","family":"monetary observations"},"economics[].currency_price_basis":{"state":"unresolved","review_state":"not_assessed","coverage":"partial","family":"monetary observations"},"economics[].gross_net":{"state":"unresolved","review_state":"not_assessed","coverage":"partial","family":"monetary observations"},"economics[].measure_type":{"state":"unresolved","review_state":"not_assessed","coverage":"partial","family":"monetary observations"},"economics[].period_unit":{"state":"unresolved","review_state":"not_assessed","coverage":"partial","family":"monetary observations"},"economics[].realization":{"state":"unresolved","review_state":"not_assessed","coverage":"partial","family":"monetary observations"},"economics[].recipient_payer":{"state":"unresolved","review_state":"not_assessed","coverage":"partial","family":"monetary observations"},"economics[].return_payback":{"state":"unresolved","review_state":"not_assessed","coverage":"partial","family":"monetary observations"},"economics[].valuation_method":{"state":"unresolved","review_state":"not_assessed","coverage":"partial","family":"monetary observations"},"evaluations[].ai_uplift":{"state":"unresolved","review_state":"not_assessed","coverage":"partial","family":"evaluations"},"evaluations[].comparator":{"state":"unresolved","review_state":"not_assessed","coverage":"partial","family":"evaluations"},"evaluations[].design":{"state":"unresolved","review_state":"not_assessed","coverage":"partial","family":"evaluations"},"evaluations[].effect":{"state":"unresolved","review_state":"not_assessed","coverage":"partial","family":"evaluations"},"evaluations[].exposure":{"state":"unresolved","review_state":"not_assessed","coverage":"partial","family":"evaluations"},"evaluations[].id":{"state":"unresolved","review_state":"not_assessed","coverage":"partial","family":"evaluations"},"evaluations[].measurement_method":{"state":"unresolved","review_state":"not_assessed","coverage":"partial","family":"evaluations"},"evaluations[].metric":{"state":"unresolved","review_state":"not_assessed","coverage":"partial","family":"evaluations"},"evaluations[].reproduction_materials":{"state":"unresolved","review_state":"not_assessed","coverage":"partial","family":"evaluations"},"evaluations[].setting_sample":{"state":"unresolved","review_state":"not_assessed","coverage":"partial","family":"evaluations"},"evaluations[].uncertainty_limitations":{"state":"unresolved","review_state":"not_assessed","coverage":"partial","family":"evaluations"},"external_references[]":{"state":"unresolved","provenance":{"kind":"curatorial","basis":"Editorial assignment.","record_fields":[],"source_refs":[]},"reason":"This is an explicit unresolved assignment: the sources on record do not settle this question.","review_state":"not_assessed","coverage":"partial"},"follow_up[]":{"state":"unresolved","provenance":{"kind":"curatorial","basis":"Editorial assignment.","record_fields":[],"source_refs":[]},"reason":"This is an explicit unresolved assignment: the sources on record do not settle this question.","review_state":"not_assessed","coverage":"partial"},"human_oversight":{"state":"unresolved","provenance":{"kind":"curatorial","basis":"Editorial assignment.","record_fields":[],"source_refs":[]},"reason":"This is an explicit unresolved assignment: the sources on record do not settle this question.","review_state":"not_assessed","coverage":"partial"},"id":{"state":"value_present","value":{"kind":"record_fields","fields":{"id":"AIBD-1031"}},"provenance":{"kind":"legacy_record","basis":"Editorial assignment.","record_fields":["id"],"source_refs":[]},"review_state":"record_mapped","coverage":"partial"},"infrastructure_dependencies[]":{"state":"unresolved","provenance":{"kind":"curatorial","basis":"Editorial assignment.","record_fields":[],"source_refs":[]},"reason":"This is an explicit unresolved assignment: the sources on record do not settle this question.","review_state":"not_assessed","coverage":"partial"},"outcomes[].beneficiaries":{"state":"unresolved","provenance":{"kind":"curatorial","basis":"Editorial assignment.","record_fields":[],"source_refs":[]},"reason":"This is an explicit unresolved assignment: the sources on record do not settle this question.","review_state":"not_assessed","coverage":"partial"},"outcomes[].beneficiary_count":{"state":"unresolved","provenance":{"kind":"curatorial","basis":"Editorial assignment.","record_fields":[],"source_refs":[]},"reason":"This is an explicit unresolved assignment: the sources on record do not settle this question.","review_state":"not_assessed","coverage":"partial"},"outcomes[].benefit_type":{"state":"value_present","value":{"kind":"record_fields","fields":{"benefit_type":"Improved Task Accuracy"}},"provenance":{"kind":"legacy_record","basis":"Editorial assignment.","record_fields":["benefit_type"],"source_refs":[]},"review_state":"record_mapped","coverage":"partial"},"outcomes[].claim":{"state":"value_present","value":{"kind":"record_fields","fields":{"account":"The Evaluation Service of the Canary Islands Health Service (SESCS) assessed wearable devices for fall risk and fall detection in people aged 65 and over. It published the assessment for Spain's national health system on 6 August 2026. Its review found 16 studies but none that measured falls, injuries or quality of life. Only one study gave the data to compute sensitivity and specificity. In that Spanish study of 157 people over 70, the G-STRIDE sensor had a sensitivity of 77% and a specificity of 83%. The Timed Up and Go test had 51% and 83%. A gait-speed test had 75% and 79%. SESCS describes G-STRIDE as a shoe-mounted inertial sensor that analyzes gait with a machine-learning algorithm. It concludes that the results suggest that G-STRIDE detects fall risk better than the Timed Up and Go test and much like the gait-speed test. The study classified people by falls in the past 12 months. SESCS rated the evidence as very low in certainty. It could not recommend the devices and suggested their use in research or pilot projects.","benefit_type":"Improved Task Accuracy"}},"provenance":{"kind":"legacy_record","basis":"Editorial assignment.","record_fields":["account","benefit_type"],"source_refs":[]},"review_state":"record_mapped","coverage":"partial"},"outcomes[].distribution":{"state":"unresolved","provenance":{"kind":"curatorial","basis":"Editorial assignment.","record_fields":[],"source_refs":[]},"reason":"This is an explicit unresolved assignment: the sources on record do not settle this question.","review_state":"not_assessed","coverage":"partial"},"outcomes[].duration":{"state":"unresolved","provenance":{"kind":"curatorial","basis":"Editorial assignment.","record_fields":[],"source_refs":[]},"reason":"This is an explicit unresolved assignment: the sources on record do not settle this question.","review_state":"not_assessed","coverage":"partial"},"outcomes[].id":{"state":"unresolved","provenance":{"kind":"curatorial","basis":"Editorial assignment.","record_fields":[],"source_refs":[]},"reason":"This is an explicit unresolved assignment: the sources on record do not settle this question.","review_state":"not_assessed","coverage":"partial"},"outcomes[].stakeholder_group":{"state":"value_present","value":{"kind":"record_fields","fields":{"ratings":[{"group":"organizational","reason":"The gain over the Timed Up and Go test comes from one research study that classified past fallers, with no measure of falls prevented."}]}},"provenance":{"kind":"legacy_record","basis":"Editorial assignment.","record_fields":["ratings"],"source_refs":[]},"review_state":"record_mapped","coverage":"partial"},"outcomes[].status":{"state":"value_present","value":{"kind":"record_fields","fields":{"benefit_status":"mixed"}},"provenance":{"kind":"legacy_record","basis":"Editorial assignment.","record_fields":["benefit_status"],"source_refs":[]},"review_state":"record_mapped","coverage":"partial"},"outcomes[].tradeoffs":{"state":"unresolved","provenance":{"kind":"curatorial","basis":"Editorial assignment.","record_fields":[],"source_refs":[]},"reason":"This is an explicit unresolved assignment: the sources on record do not settle this question.","review_state":"not_assessed","coverage":"partial"},"outcomes[].unexpected":{"state":"value_present","value":{"kind":"record_fields","fields":{"unexpected":false,"purpose":"An academic consortium of Universidad Rey Juan Carlos and collaborators built G-STRIDE to analyze gait with a shoe-mounted inertial sensor and estimate the risk of falling in older people."}},"provenance":{"kind":"legacy_record","basis":"Editorial assignment.","record_fields":["unexpected","purpose"],"source_refs":[]},"review_state":"record_mapped","coverage":"partial"},"places[]":{"state":"value_present","value":{"kind":"record_fields","fields":{"location":"Spain","developer_country":"Spain","deployer_country":"Spain"}},"provenance":{"kind":"legacy_record","basis":"Editorial assignment.","record_fields":["location","developer_country","deployer_country"],"source_refs":[]},"review_state":"record_mapped","coverage":"partial"},"purpose":{"state":"value_present","value":{"kind":"record_fields","fields":{"purpose":"An academic consortium of Universidad Rey Juan Carlos and collaborators built G-STRIDE to analyze gait with a shoe-mounted inertial sensor and estimate the risk of falling in older people."}},"provenance":{"kind":"legacy_record","basis":"Editorial assignment.","record_fields":["purpose"],"source_refs":[]},"review_state":"record_mapped","coverage":"partial"},"record_dates":{"state":"value_present","value":{"kind":"record_fields","fields":{"date_observed":"2026-08-06","date_basis":"source publication","date_precision":"day","revisions":[{"date":"2026-11-01","change":"Added the record from the finding of the Evaluation Service of the Canary Islands Health Service (SESCS).","source":"https://sescs.es/wp-content/uploads/2026/08/114_2025_SESCS_Telemonitorizacion_Caidas_DEF_NIPO.pdf","editor":"AIBD"}]}},"provenance":{"kind":"legacy_record","basis":"Editorial assignment.","record_fields":["date_observed","date_basis","date_precision","revisions"],"source_refs":[]},"review_state":"record_mapped","coverage":"partial"},"replication.attempts[]":{"state":"unresolved","provenance":{"kind":"curatorial","basis":"Editorial assignment.","record_fields":[],"source_refs":[]},"reason":"This is an explicit unresolved assignment: the sources on record do not settle this question.","review_state":"not_assessed","coverage":"partial"},"replication.conditions":{"state":"value_present","value":{"kind":"record_fields","fields":{"replication":{"produced_by":["An inertial measurement unit on the foot recorded gait parameters while people walked without restrictions.","A machine-learning algorithm classified fall risk from the gait parameters."],"must_hold":["A clinician must interpret the result in a consultation, as with the conventional tests.","Sensor and software costs must stay low, since the cost per extra correct case was lowest with large-scale purchase."],"echoes":[]}}},"provenance":{"kind":"legacy_record","basis":"Editorial assignment.","record_fields":["replication"],"source_refs":[]},"review_state":"record_mapped","coverage":"partial"},"revision_history[]":{"state":"value_present","value":{"kind":"record_fields","fields":{"revisions":[{"date":"2026-11-01","change":"Added the record from the finding of the Evaluation Service of the Canary Islands Health Service (SESCS).","source":"https://sescs.es/wp-content/uploads/2026/08/114_2025_SESCS_Telemonitorizacion_Caidas_DEF_NIPO.pdf","editor":"AIBD"}]}},"provenance":{"kind":"legacy_record","basis":"Editorial assignment.","record_fields":["revisions"],"source_refs":[]},"review_state":"record_mapped","coverage":"partial"},"sector":{"state":"value_present","value":{"kind":"record_fields","fields":{"domain":"Healthcare (fall prevention in older people)"}},"provenance":{"kind":"legacy_record","basis":"Editorial assignment.","record_fields":["domain"],"source_refs":[]},"review_state":"record_mapped","coverage":"partial"},"sources[].access_version":{"state":"unresolved","provenance":{"kind":"curatorial","basis":"Editorial assignment.","record_fields":[],"source_refs":[]},"reason":"This is an explicit unresolved assignment: the sources on record do not settle this question.","review_state":"not_assessed","coverage":"partial"},"sources[].citation_locator":{"state":"value_present","value":{"kind":"record_fields","fields":{"evidence":[{"kind":"assessment report","citation":"Servicio de Evaluación del Servicio Canario de la Salud (SESCS). Sistemas de telemonitorización portátiles para la evaluación del riesgo de caídas y la detección de caídas en personas de 65 o más años. Informes de Evaluación de Tecnologías Sanitarias, Ministerio de Sanidad, 2026. Resumen ejecutivo (Resultados, Coste-efectividad y análisis económico).","title":"Sistemas de telemonitorización portátiles para la evaluación del riesgo de caídas y la detección de caídas en personas de 65 o más años","url":"https://sescs.es/wp-content/uploads/2026/08/114_2025_SESCS_Telemonitorizacion_Caidas_DEF_NIPO.pdf","date":"2026-08-06","written_by":"independent","quote":"Los resultados sugieren que G-STRIDE mejora la detección del riesgo de caída respecto al test TUG, aunque su rendimiento es muy similar al test GS."}]}},"provenance":{"kind":"legacy_record","basis":"Editorial assignment.","record_fields":["evidence"],"source_refs":[]},"review_state":"record_mapped","coverage":"partial"},"sources[].id":{"state":"value_present","value":{"kind":"identifiers","identifiers":["AIBD-1031-source-0001"],"role":"Existing evidence items, in their original order"},"provenance":{"kind":"curatorial","basis":"Editorial assignment.","record_fields":[],"source_refs":[]},"review_state":"record_mapped","coverage":"complete"},"sources[].language":{"state":"unresolved","provenance":{"kind":"curatorial","basis":"Editorial assignment.","record_fields":[],"source_refs":[]},"reason":"This is an explicit unresolved assignment: the sources on record do not settle this question.","review_state":"not_assessed","coverage":"partial"},"sources[].relationship":{"state":"value_present","value":{"kind":"record_fields","fields":{"evidence":[{"kind":"assessment report","citation":"Servicio de Evaluación del Servicio Canario de la Salud (SESCS). Sistemas de telemonitorización portátiles para la evaluación del riesgo de caídas y la detección de caídas en personas de 65 o más años. Informes de Evaluación de Tecnologías Sanitarias, Ministerio de Sanidad, 2026. Resumen ejecutivo (Resultados, Coste-efectividad y análisis económico).","title":"Sistemas de telemonitorización portátiles para la evaluación del riesgo de caídas y la detección de caídas en personas de 65 o más años","url":"https://sescs.es/wp-content/uploads/2026/08/114_2025_SESCS_Telemonitorizacion_Caidas_DEF_NIPO.pdf","date":"2026-08-06","written_by":"independent","quote":"Los resultados sugieren que G-STRIDE mejora la detección del riesgo de caída respecto al test TUG, aunque su rendimiento es muy similar al test GS."}]}},"provenance":{"kind":"legacy_record","basis":"Editorial assignment.","record_fields":["evidence"],"source_refs":[]},"review_state":"record_mapped","coverage":"partial"},"sources[].upstream_sources":{"state":"unresolved","provenance":{"kind":"curatorial","basis":"Editorial assignment.","record_fields":[],"source_refs":[]},"reason":"This is an explicit unresolved assignment: the sources on record do not settle this question.","review_state":"not_assessed","coverage":"partial"},"strategic_assessments[]":{"state":"unresolved","review_state":"not_assessed","coverage":"partial","family":"strategic assessments"},"summary":{"state":"value_present","value":{"kind":"record_fields","fields":{"summary":"SESCS found that, in one study, the G-STRIDE gait sensor identified more older people at risk of falling than the Timed Up and Go test and about as many as a gait-speed test.","account":"The Evaluation Service of the Canary Islands Health Service (SESCS) assessed wearable devices for fall risk and fall detection in people aged 65 and over. It published the assessment for Spain's national health system on 6 August 2026. Its review found 16 studies but none that measured falls, injuries or quality of life. Only one study gave the data to compute sensitivity and specificity. In that Spanish study of 157 people over 70, the G-STRIDE sensor had a sensitivity of 77% and a specificity of 83%. The Timed Up and Go test had 51% and 83%. A gait-speed test had 75% and 79%. SESCS describes G-STRIDE as a shoe-mounted inertial sensor that analyzes gait with a machine-learning algorithm. It concludes that the results suggest that G-STRIDE detects fall risk better than the Timed Up and Go test and much like the gait-speed test. The study classified people by falls in the past 12 months. SESCS rated the evidence as very low in certainty. It could not recommend the devices and suggested their use in research or pilot projects."}},"provenance":{"kind":"legacy_record","basis":"Editorial assignment.","record_fields":["summary","account"],"source_refs":[]},"review_state":"record_mapped","coverage":"partial"},"systems[].autonomy_permissions":{"state":"unresolved","provenance":{"kind":"curatorial","basis":"Editorial assignment.","record_fields":[],"source_refs":[]},"reason":"This is an explicit unresolved assignment: the sources on record do not settle this question.","review_state":"not_assessed","coverage":"partial"},"systems[].component_role":{"state":"unresolved","provenance":{"kind":"curatorial","basis":"Editorial assignment.","record_fields":[],"source_refs":[]},"reason":"This is an explicit unresolved assignment: the sources on record do not settle this question.","review_state":"not_assessed","coverage":"partial"},"systems[].deployment_configuration":{"state":"unresolved","provenance":{"kind":"curatorial","basis":"Editorial assignment.","record_fields":[],"source_refs":[]},"reason":"This is an explicit unresolved assignment: the sources on record do not settle this question.","review_state":"not_assessed","coverage":"partial"},"systems[].inputs_outputs":{"state":"unresolved","provenance":{"kind":"curatorial","basis":"Editorial assignment.","record_fields":[],"source_refs":[]},"reason":"This is an explicit unresolved assignment: the sources on record do not settle this question.","review_state":"not_assessed","coverage":"partial"},"systems[].model_name":{"state":"unresolved","provenance":{"kind":"curatorial","basis":"Editorial assignment.","record_fields":[],"source_refs":[]},"reason":"This is an explicit unresolved assignment: the sources on record do not settle this question.","review_state":"not_assessed","coverage":"partial"},"systems[].model_version":{"state":"unresolved","provenance":{"kind":"curatorial","basis":"Editorial assignment.","record_fields":[],"source_refs":[]},"reason":"This is an explicit unresolved assignment: the sources on record do not settle this question.","review_state":"not_assessed","coverage":"partial"},"systems[].product_name":{"state":"value_present","value":{"kind":"record_fields","fields":{"system":"G-STRIDE (shoe-mounted inertial sensor with a machine-learning gait algorithm)"}},"provenance":{"kind":"legacy_record","basis":"Editorial assignment.","record_fields":["system"],"source_refs":[]},"review_state":"record_mapped","coverage":"partial"},"systems[].provider":{"state":"unresolved","provenance":{"kind":"curatorial","basis":"Editorial assignment.","record_fields":[],"source_refs":[]},"reason":"This is an explicit unresolved assignment: the sources on record do not settle this question.","review_state":"not_assessed","coverage":"partial"},"taxonomy_crosswalks[]":{"state":"unresolved","provenance":{"kind":"curatorial","basis":"Editorial assignment.","record_fields":[],"source_refs":[]},"reason":"This is an explicit unresolved assignment: the sources on record do not settle this question.","review_state":"not_assessed","coverage":"partial"},"title":{"state":"value_present","value":{"kind":"record_fields","fields":{"title":"Canary Islands assessment service SESCS finds that the G-STRIDE gait sensor may detect fall risk better than the Timed Up and Go test","headline":"*G-STRIDE* spots more older people at risk of falls than the TUG test"}},"provenance":{"kind":"legacy_record","basis":"Editorial assignment.","record_fields":["title","headline"],"source_refs":[]},"review_state":"record_mapped","coverage":"partial"},"version":{"state":"value_present","value":{"kind":"identifiers","identifiers":["3fac9b2cd450f5c524a5510ad94734cb10cf9b1f5c49d94a7735fb3767fc6b4f"],"role":"SHA-256 of the complete public record and its assessments"},"provenance":{"kind":"curatorial","basis":"SHA-256 of the public record, its assessments, and its displayed answers and notes, excluding its own identifier and the derived coverage summary.","record_fields":[],"source_refs":[]},"review_state":"record_mapped","coverage":"complete"},"workflow":{"state":"value_present","value":{"kind":"record_fields","fields":{"account":"The Evaluation Service of the Canary Islands Health Service (SESCS) assessed wearable devices for fall risk and fall detection in people aged 65 and over. It published the assessment for Spain's national health system on 6 August 2026. Its review found 16 studies but none that measured falls, injuries or quality of life. Only one study gave the data to compute sensitivity and specificity. In that Spanish study of 157 people over 70, the G-STRIDE sensor had a sensitivity of 77% and a specificity of 83%. The Timed Up and Go test had 51% and 83%. A gait-speed test had 75% and 79%. SESCS describes G-STRIDE as a shoe-mounted inertial sensor that analyzes gait with a machine-learning algorithm. It concludes that the results suggest that G-STRIDE detects fall risk better than the Timed Up and Go test and much like the gait-speed test. The study classified people by falls in the past 12 months. SESCS rated the evidence as very low in certainty. It could not recommend the devices and suggested their use in research or pilot projects.","replication":{"produced_by":["An inertial measurement unit on the foot recorded gait parameters while people walked without restrictions.","A machine-learning algorithm classified fall risk from the gait parameters."],"must_hold":["A clinician must interpret the result in a consultation, as with the conventional tests.","Sensor and software costs must stay low, since the cost per extra correct case was lowest with large-scale purchase."],"echoes":[]}}},"provenance":{"kind":"legacy_record","basis":"Editorial assignment.","record_fields":["account","replication"],"source_refs":[]},"review_state":"record_mapped","coverage":"partial"}},"summary":{"total":91,"addressed":91,"known":28,"reviewed":2,"reviewed_complete":2,"reviewed_partial":0,"record_mapped":26,"review_not_assessed":63,"not_assessed":0,"complete":false,"by_state":{"value_present":28,"not_reported":0,"unresolved":63,"not_assessed":0,"not_applicable":0,"withheld":0,"source_unavailable":0}}},"presentation_version":"b1ec8249a69dd17790f15e8ffa8dd3bca74dc1d9076b973d30e1ee863e300543"}