{"id":"AIBD-0733","title":"WHO finds that six AI chest X-ray programs, including qXR v4.0 and CAD4TB v7, meet its standards for TB screening of people aged 15 and older","summary":"WHO found that six of eight AI programs tested on 3,915 archived chest X-rays met its accuracy standards for TB screening of people aged 15 and older.","date_observed":"2025-06-10","date_basis":"source publication","date_precision":"day","benefit_status":"reported","source_collection":null,"source_entry":null,"account":"Computer-aided detection software reads a digital chest X-ray and returns a score for the likelihood of tuberculosis. WHO says that such systems are trained on thousands or millions of images using artificial intelligence. In 2024, WHO invited makers to submit products for an independent evaluation on a validation platform run by FIND in Geneva. Eight products met regulatory requirements. They were tested on two image libraries, 1,663 images for community screening and 2,252 for facility screening, which the makers had not used for training. Microbiological test results served as the reference, and radiologists read the same images for comparison. To pass, a product needed at least 40% specificity at 90% sensitivity and a high partial area under the curve in both libraries. WHO's Technical Advisory Group reviewed the results in January 2025. In a policy statement published on 10 June 2025, WHO found that six products met its standards. In community screening their specificity at 90% sensitivity ranged from 45% to 56%. Accuracy was generally lower for people living with HIV, women and people with a history of TB.","category":"aggregate result","unexpected":false,"purpose":"Delft Imaging, VinBrain, DeepTek, Infervision, Lunit and Qure.ai built their programs to read digital chest X-rays and score the likelihood of tuberculosis so that screening programs can decide who needs further testing.","system":"CAD4TB, DrAid for TB screening, Genki Edge, InferRead DR Chest, Insight CXR and qXR (computer-aided detection software for TB)","model_version":"CAD4TB v7, DrAid for TB screening v2.4.4–6, Genki Edge v3.4.2, InferRead DR Chest v1.0.1.1, Insight CXR v3.1 and qXR v4.0","developer":"Delft Imaging (CAD4TB), VinBrain (DrAid), DeepTek (Genki Edge), Infervision (InferRead DR Chest), Lunit (Insight CXR) and Qure.ai (qXR)","deployer":"FIND's independent validation platform in Geneva, for WHO (archived chest X-ray libraries)","location":"Global (evaluation in Geneva, Switzerland)","domain":"Healthcare (tuberculosis screening)","segment":null,"benefit_type":"Improved Task Accuracy","work_activities":["4.A.1.b.1"],"taxonomy_note":null,"registry_ids":{},"ratings":[{"group":"organizational","depth":null,"reason":"The products were tested on archived images, so no screening program's results with them are reported."}],"ai_role":{"status":"assessed","statement":"The software read each image alone, and its accuracy was measured against microbiological results, with radiologists' reads of the same images as the comparator. The makers had not used the images for training."},"evidence":[{"kind":"assessment report","citation":"World Health Organization, Global Programme on Tuberculosis & Lung Health. Use of computer-aided detection software for tuberculosis screening: WHO policy statement. Geneva, 10 June 2025. ISBN 978-92-4-011037-3.","title":"Use of computer-aided detection software for tuberculosis screening: WHO policy statement","url":"https://iris.who.int/server/api/core/bitstreams/0d7e8935-879e-4ab6-8c07-e41f7be1465b/content","date":"2025-06-10","written_by":"independent","quote":"CAD4TB v7, DrAid for TB screening v2.4.4–6, Genki Edge v3.4.2, InferRead DR Chest v1.0.1.1, Insight CXR v3.1 and qXR v4.0 met the WHO performance standards for the use of CAD in community-based and facility-based screening for pulmonary TB in people aged 15 years and older"}],"replication":{"produced_by":["Each product was set to a threshold giving 90% sensitivity against microbiological results.","The image libraries came from earlier screening studies and were never shared with the makers."],"must_hold":["Programs calibrate the threshold for their own setting, hardware and population.","Programs have laboratory capacity for the confirmatory tests that positive screens require."],"echoes":[]},"paired_incident":null,"headline":"Six AI chest X-ray readers meet WHO's standard for TB screening","metric":"ranged from 45% to 56%","comparator":{"against":"WHO's minimum of 40% specificity at 90% sensitivity","before":"40%","after":"45% to 56% for the six products in community screening","basis":"measured","quote":"The specificity of CAD software products ranged between 26% and 56%. For one product – Radify Chest v3.8.0 – the specificity at 90% sensitivity did not reach 40%.","url":"https://iris.who.int/server/api/core/bitstreams/0d7e8935-879e-4ab6-8c07-e41f7be1465b/content"},"later":null,"value":null,"developer_country":"Multiple: Netherlands, Vietnam, India, China, South Korea","deployer_country":"Switzerland","unrated_because":"WHO measured accuracy on archived images and reports no outcome of screening with the products.","brief_notes":null,"brief":null,"revisions":[{"date":"2026-10-31","change":"Added the record from the finding of WHO.","source":"https://iris.who.int/server/api/core/bitstreams/0d7e8935-879e-4ab6-8c07-e41f7be1465b/content","editor":"AIBD"}],"confidence":"found","shade":"high","claim_status":"Found by an assessment body","claim_basis":"Sources: one independent assessment report. The cited evidence includes no independent study or report.","case_jurisdictions":["Switzerland"],"profile":{"case_id":"AIBD-0733","catalog_version":"1.0.0","source_fingerprint":"e681bcc32fa87ffa096ffc6f8ff7358e8bbfc5d765927781ef91946a17d3ca6b","collection_refs":[{"collection_id":"aibd","version":"2026-10-31"},{"collection_id":"querist","version":"2026-10-25"}],"source_ids":["AIBD-0733-source-0001"],"families":{"strategic assessments":{"state":"unresolved","provenance":{"kind":"curatorial","basis":"Editorial assignment.","record_fields":[],"source_refs":[]},"reason":"Editorial assignment.","review_state":"not_assessed","coverage":"partial","record_ids":[]},"evaluations":{"state":"unresolved","provenance":{"kind":"curatorial","basis":"Editorial assignment.","record_fields":[],"source_refs":[]},"reason":"Editorial assignment.","review_state":"not_assessed","coverage":"partial","record_ids":[]},"monetary observations":{"state":"unresolved","provenance":{"kind":"curatorial","basis":"Editorial assignment.","record_fields":[],"source_refs":[]},"reason":"Editorial assignment.","review_state":"not_assessed","coverage":"partial","record_ids":[]},"institutional decisions":{"state":"unresolved","provenance":{"kind":"curatorial","basis":"Editorial assignment.","record_fields":[],"source_refs":[]},"reason":"Editorial assignment.","review_state":"not_assessed","coverage":"partial","record_ids":[]}},"fields":{"actors[]":{"state":"value_present","value":{"kind":"record_fields","fields":{"developer":"Delft Imaging (CAD4TB), VinBrain (DrAid), DeepTek (Genki Edge), Infervision (InferRead DR Chest), Lunit (Insight CXR) and Qure.ai (qXR)","deployer":"FIND's independent validation platform in Geneva, for WHO (archived chest X-ray libraries)"}},"provenance":{"kind":"legacy_record","basis":"Editorial assignment.","record_fields":["developer","deployer"],"source_refs":[]},"review_state":"record_mapped","coverage":"partial"},"actors[].ownership_control":{"state":"unresolved","provenance":{"kind":"curatorial","basis":"Editorial assignment.","record_fields":[],"source_refs":[]},"reason":"This is an explicit unresolved assignment: the sources on record do not settle this question.","review_state":"not_assessed","coverage":"partial"},"ai_contribution":{"state":"value_present","value":{"kind":"record_fields","fields":{"ai_role":{"status":"assessed","statement":"The software read each image alone, and its accuracy was measured against microbiological results, with radiologists' reads of the same images as the comparator. The makers had not used the images for training."}}},"provenance":{"kind":"legacy_record","basis":"Editorial assignment.","record_fields":["ai_role"],"source_refs":[]},"review_state":"record_mapped","coverage":"partial"},"ai_relationship":{"state":"value_present","value":{"kind":"record_fields","fields":{"category":"aggregate result","ai_role":{"status":"assessed","statement":"The software read each image alone, and its accuracy was measured against microbiological results, with radiologists' reads of the same images as the comparator. The makers had not used the images for training."}}},"provenance":{"kind":"legacy_record","basis":"Editorial assignment.","record_fields":["ai_role","category"],"source_refs":[]},"review_state":"record_mapped","coverage":"partial"},"assessments[].alternatives":{"state":"unresolved","provenance":{"kind":"curatorial","basis":"Editorial assignment.","record_fields":[],"source_refs":[]},"reason":"This is an explicit unresolved assignment: the sources on record do not settle this question.","review_state":"not_assessed","coverage":"partial"},"assessments[].assessor_date":{"state":"unresolved","provenance":{"kind":"curatorial","basis":"Editorial assignment.","record_fields":[],"source_refs":[]},"reason":"This is an explicit unresolved assignment: the sources on record do not settle this question.","review_state":"not_assessed","coverage":"partial"},"assessments[].depth":{"state":"unresolved","provenance":{"kind":"curatorial","basis":"Editorial assignment.","record_fields":[],"source_refs":[]},"reason":"This is an explicit unresolved assignment: the sources on record do not settle this question.","review_state":"not_assessed","coverage":"partial"},"assessments[].display_confidence":{"state":"value_present","value":{"kind":"structured","data":{"confidence":"high","evidence_word":"found","claim_status":"Found by an assessment body","basis":"Sources: one independent assessment report. The cited evidence includes no independent study or report.","interpretation":"Existing display rule preserved; the richer outcome-specific confidence question still needs review."}},"provenance":{"kind":"curatorial","basis":"Editorial assignment.","record_fields":[],"source_refs":[]},"review_state":"record_mapped","coverage":"partial"},"assessments[].rationale":{"state":"value_present","value":{"kind":"record_fields","fields":{"ratings":[{"group":"organizational","reason":"The products were tested on archived images, so no screening program's results with them are reported."}]}},"provenance":{"kind":"legacy_record","basis":"Editorial assignment.","record_fields":["ratings"],"source_refs":[]},"review_state":"record_mapped","coverage":"partial"},"assessments[].rubric_version":{"state":"unresolved","provenance":{"kind":"curatorial","basis":"Editorial assignment.","record_fields":[],"source_refs":[]},"reason":"This is an explicit unresolved assignment: the sources on record do not settle this question.","review_state":"not_assessed","coverage":"partial"},"assessments[].status_basis":{"state":"unresolved","provenance":{"kind":"curatorial","basis":"Editorial assignment.","record_fields":[],"source_refs":[]},"reason":"This is an explicit unresolved assignment: the sources on record do not settle this question.","review_state":"not_assessed","coverage":"partial"},"assessments[].target":{"state":"value_present","value":{"kind":"record_fields","fields":{"ratings":[{"group":"organizational","reason":"The products were tested on archived images, so no screening program's results with them are reported."}]}},"provenance":{"kind":"legacy_record","basis":"Editorial assignment.","record_fields":["ratings"],"source_refs":[]},"review_state":"record_mapped","coverage":"partial"},"availability":{"state":"unresolved","provenance":{"kind":"curatorial","basis":"Editorial assignment.","record_fields":[],"source_refs":[]},"reason":"This is an explicit unresolved assignment: the sources on record do not settle this question.","review_state":"not_assessed","coverage":"partial"},"brief":{"state":"unresolved","provenance":{"kind":"curatorial","basis":"Editorial assignment.","record_fields":[],"source_refs":[]},"reason":"This is an explicit unresolved assignment: the sources on record do not settle this question.","review_state":"not_assessed","coverage":"partial"},"case_kind":{"state":"unresolved","provenance":{"kind":"curatorial","basis":"Editorial assignment.","record_fields":[],"source_refs":[]},"reason":"This is an explicit unresolved assignment: the sources on record do not settle this question.","review_state":"not_assessed","coverage":"partial"},"case_links[]":{"state":"unresolved","provenance":{"kind":"curatorial","basis":"Editorial assignment.","record_fields":[],"source_refs":[]},"reason":"This is an explicit unresolved assignment: the sources on record do not settle this question.","review_state":"not_assessed","coverage":"partial"},"claims[].evidence_links":{"state":"unresolved","provenance":{"kind":"curatorial","basis":"Editorial assignment.","record_fields":[],"source_refs":[]},"reason":"This is an explicit unresolved assignment: the sources on record do not settle this question.","review_state":"not_assessed","coverage":"partial"},"claims[].evidence_strength":{"state":"value_present","value":{"kind":"structured","data":{"answer":"Found by an assessment body. Sources: one independent assessment report. The cited evidence includes no independent study or report.","evidence_word":"found","claim_status":"Found by an assessment body"}},"provenance":{"kind":"source_review","basis":"Editorial assignment.","record_fields":["evidence"],"source_refs":["AIBD-0733-source-0001"],"reviewed_on":"2026-10-29"},"review_state":"reviewed","coverage":"complete","display":{"answer":"Found by an assessment body. Sources: one independent assessment report. The cited evidence includes no independent study or report.","inferred":false}},"claims[].missingness":{"state":"unresolved","provenance":{"kind":"curatorial","basis":"Editorial assignment.","record_fields":[],"source_refs":[]},"reason":"This is an explicit unresolved assignment: the sources on record do not settle this question.","review_state":"not_assessed","coverage":"partial"},"claims[].proposition":{"state":"unresolved","provenance":{"kind":"curatorial","basis":"Editorial assignment.","record_fields":[],"source_refs":[]},"reason":"This is an explicit unresolved assignment: the sources on record do not settle this question.","review_state":"not_assessed","coverage":"partial"},"cross_border_flows[]":{"state":"unresolved","provenance":{"kind":"curatorial","basis":"Editorial assignment.","record_fields":[],"source_refs":[]},"reason":"This is an explicit unresolved assignment: the sources on record do not settle this question.","review_state":"not_assessed","coverage":"partial"},"decisions[].action_outcome":{"state":"unresolved","review_state":"not_assessed","coverage":"partial","family":"institutional decisions"},"decisions[].grounds":{"state":"unresolved","review_state":"not_assessed","coverage":"partial","family":"institutional decisions"},"decisions[].institution_task":{"state":"unresolved","review_state":"not_assessed","coverage":"partial","family":"institutional decisions"},"decisions[].success_criterion":{"state":"unresolved","review_state":"not_assessed","coverage":"partial","family":"institutional decisions"},"deployment_stage":{"state":"value_present","value":{"kind":"structured","data":{"classification":"laboratory test","answer":"The system was tested in a laboratory. WHO's evaluation ran the products on archived chest X-ray libraries on FIND's validation platform.","scope":"The period and activity described in the case."}},"provenance":{"kind":"source_review","basis":"Editorial assignment.","record_fields":["account","evidence"],"source_refs":["AIBD-0733-source-0001"],"reviewed_on":"2026-10-29"},"review_state":"reviewed","coverage":"complete","display":{"answer":"The system was tested in a laboratory. WHO's evaluation ran the products on archived chest X-ray libraries on FIND's validation platform.","inferred":false}},"economics[].amount":{"state":"unresolved","review_state":"not_assessed","coverage":"partial","family":"monetary observations"},"economics[].costs[]":{"state":"unresolved","review_state":"not_assessed","coverage":"partial","family":"monetary observations"},"economics[].counterfactual":{"state":"unresolved","review_state":"not_assessed","coverage":"partial","family":"monetary observations"},"economics[].currency_price_basis":{"state":"unresolved","review_state":"not_assessed","coverage":"partial","family":"monetary observations"},"economics[].gross_net":{"state":"unresolved","review_state":"not_assessed","coverage":"partial","family":"monetary observations"},"economics[].measure_type":{"state":"unresolved","review_state":"not_assessed","coverage":"partial","family":"monetary observations"},"economics[].period_unit":{"state":"unresolved","review_state":"not_assessed","coverage":"partial","family":"monetary observations"},"economics[].realization":{"state":"unresolved","review_state":"not_assessed","coverage":"partial","family":"monetary observations"},"economics[].recipient_payer":{"state":"unresolved","review_state":"not_assessed","coverage":"partial","family":"monetary observations"},"economics[].return_payback":{"state":"unresolved","review_state":"not_assessed","coverage":"partial","family":"monetary observations"},"economics[].valuation_method":{"state":"unresolved","review_state":"not_assessed","coverage":"partial","family":"monetary observations"},"evaluations[].ai_uplift":{"state":"unresolved","review_state":"not_assessed","coverage":"partial","family":"evaluations"},"evaluations[].comparator":{"state":"unresolved","review_state":"not_assessed","coverage":"partial","family":"evaluations"},"evaluations[].design":{"state":"unresolved","review_state":"not_assessed","coverage":"partial","family":"evaluations"},"evaluations[].effect":{"state":"unresolved","review_state":"not_assessed","coverage":"partial","family":"evaluations"},"evaluations[].exposure":{"state":"unresolved","review_state":"not_assessed","coverage":"partial","family":"evaluations"},"evaluations[].id":{"state":"unresolved","review_state":"not_assessed","coverage":"partial","family":"evaluations"},"evaluations[].measurement_method":{"state":"unresolved","review_state":"not_assessed","coverage":"partial","family":"evaluations"},"evaluations[].metric":{"state":"unresolved","review_state":"not_assessed","coverage":"partial","family":"evaluations"},"evaluations[].reproduction_materials":{"state":"unresolved","review_state":"not_assessed","coverage":"partial","family":"evaluations"},"evaluations[].setting_sample":{"state":"unresolved","review_state":"not_assessed","coverage":"partial","family":"evaluations"},"evaluations[].uncertainty_limitations":{"state":"unresolved","review_state":"not_assessed","coverage":"partial","family":"evaluations"},"external_references[]":{"state":"unresolved","provenance":{"kind":"curatorial","basis":"Editorial assignment.","record_fields":[],"source_refs":[]},"reason":"This is an explicit unresolved assignment: the sources on record do not settle this question.","review_state":"not_assessed","coverage":"partial"},"follow_up[]":{"state":"unresolved","provenance":{"kind":"curatorial","basis":"Editorial assignment.","record_fields":[],"source_refs":[]},"reason":"This is an explicit unresolved assignment: the sources on record do not settle this question.","review_state":"not_assessed","coverage":"partial"},"human_oversight":{"state":"unresolved","provenance":{"kind":"curatorial","basis":"Editorial assignment.","record_fields":[],"source_refs":[]},"reason":"This is an explicit unresolved assignment: the sources on record do not settle this question.","review_state":"not_assessed","coverage":"partial"},"id":{"state":"value_present","value":{"kind":"record_fields","fields":{"id":"AIBD-0733"}},"provenance":{"kind":"legacy_record","basis":"Editorial assignment.","record_fields":["id"],"source_refs":[]},"review_state":"record_mapped","coverage":"partial"},"infrastructure_dependencies[]":{"state":"unresolved","provenance":{"kind":"curatorial","basis":"Editorial assignment.","record_fields":[],"source_refs":[]},"reason":"This is an explicit unresolved assignment: the sources on record do not settle this question.","review_state":"not_assessed","coverage":"partial"},"outcomes[].beneficiaries":{"state":"unresolved","provenance":{"kind":"curatorial","basis":"Editorial assignment.","record_fields":[],"source_refs":[]},"reason":"This is an explicit unresolved assignment: the sources on record do not settle this question.","review_state":"not_assessed","coverage":"partial"},"outcomes[].beneficiary_count":{"state":"unresolved","provenance":{"kind":"curatorial","basis":"Editorial assignment.","record_fields":[],"source_refs":[]},"reason":"This is an explicit unresolved assignment: the sources on record do not settle this question.","review_state":"not_assessed","coverage":"partial"},"outcomes[].benefit_type":{"state":"value_present","value":{"kind":"record_fields","fields":{"benefit_type":"Improved Task Accuracy"}},"provenance":{"kind":"legacy_record","basis":"Editorial assignment.","record_fields":["benefit_type"],"source_refs":[]},"review_state":"record_mapped","coverage":"partial"},"outcomes[].claim":{"state":"value_present","value":{"kind":"record_fields","fields":{"account":"Computer-aided detection software reads a digital chest X-ray and returns a score for the likelihood of tuberculosis. WHO says that such systems are trained on thousands or millions of images using artificial intelligence. In 2024, WHO invited makers to submit products for an independent evaluation on a validation platform run by FIND in Geneva. Eight products met regulatory requirements. They were tested on two image libraries, 1,663 images for community screening and 2,252 for facility screening, which the makers had not used for training. Microbiological test results served as the reference, and radiologists read the same images for comparison. To pass, a product needed at least 40% specificity at 90% sensitivity and a high partial area under the curve in both libraries. WHO's Technical Advisory Group reviewed the results in January 2025. In a policy statement published on 10 June 2025, WHO found that six products met its standards. In community screening their specificity at 90% sensitivity ranged from 45% to 56%. Accuracy was generally lower for people living with HIV, women and people with a history of TB.","benefit_type":"Improved Task Accuracy"}},"provenance":{"kind":"legacy_record","basis":"Editorial assignment.","record_fields":["account","benefit_type"],"source_refs":[]},"review_state":"record_mapped","coverage":"partial"},"outcomes[].distribution":{"state":"unresolved","provenance":{"kind":"curatorial","basis":"Editorial assignment.","record_fields":[],"source_refs":[]},"reason":"This is an explicit unresolved assignment: the sources on record do not settle this question.","review_state":"not_assessed","coverage":"partial"},"outcomes[].duration":{"state":"unresolved","provenance":{"kind":"curatorial","basis":"Editorial assignment.","record_fields":[],"source_refs":[]},"reason":"This is an explicit unresolved assignment: the sources on record do not settle this question.","review_state":"not_assessed","coverage":"partial"},"outcomes[].id":{"state":"unresolved","provenance":{"kind":"curatorial","basis":"Editorial assignment.","record_fields":[],"source_refs":[]},"reason":"This is an explicit unresolved assignment: the sources on record do not settle this question.","review_state":"not_assessed","coverage":"partial"},"outcomes[].stakeholder_group":{"state":"value_present","value":{"kind":"record_fields","fields":{"ratings":[{"group":"organizational","reason":"The products were tested on archived images, so no screening program's results with them are reported."}]}},"provenance":{"kind":"legacy_record","basis":"Editorial assignment.","record_fields":["ratings"],"source_refs":[]},"review_state":"record_mapped","coverage":"partial"},"outcomes[].status":{"state":"value_present","value":{"kind":"record_fields","fields":{"benefit_status":"reported"}},"provenance":{"kind":"legacy_record","basis":"Editorial assignment.","record_fields":["benefit_status"],"source_refs":[]},"review_state":"record_mapped","coverage":"partial"},"outcomes[].tradeoffs":{"state":"unresolved","provenance":{"kind":"curatorial","basis":"Editorial assignment.","record_fields":[],"source_refs":[]},"reason":"This is an explicit unresolved assignment: the sources on record do not settle this question.","review_state":"not_assessed","coverage":"partial"},"outcomes[].unexpected":{"state":"value_present","value":{"kind":"record_fields","fields":{"unexpected":false,"purpose":"Delft Imaging, VinBrain, DeepTek, Infervision, Lunit and Qure.ai built their programs to read digital chest X-rays and score the likelihood of tuberculosis so that screening programs can decide who needs further testing."}},"provenance":{"kind":"legacy_record","basis":"Editorial assignment.","record_fields":["unexpected","purpose"],"source_refs":[]},"review_state":"record_mapped","coverage":"partial"},"places[]":{"state":"value_present","value":{"kind":"record_fields","fields":{"location":"Global (evaluation in Geneva, Switzerland)","developer_country":"Multiple: Netherlands, Vietnam, India, China, South Korea","deployer_country":"Switzerland"}},"provenance":{"kind":"legacy_record","basis":"Editorial assignment.","record_fields":["location","developer_country","deployer_country"],"source_refs":[]},"review_state":"record_mapped","coverage":"partial"},"purpose":{"state":"value_present","value":{"kind":"record_fields","fields":{"purpose":"Delft Imaging, VinBrain, DeepTek, Infervision, Lunit and Qure.ai built their programs to read digital chest X-rays and score the likelihood of tuberculosis so that screening programs can decide who needs further testing."}},"provenance":{"kind":"legacy_record","basis":"Editorial assignment.","record_fields":["purpose"],"source_refs":[]},"review_state":"record_mapped","coverage":"partial"},"record_dates":{"state":"value_present","value":{"kind":"record_fields","fields":{"date_observed":"2025-06-10","date_basis":"source publication","date_precision":"day","revisions":[{"date":"2026-10-31","change":"Added the record from the finding of WHO.","source":"https://iris.who.int/server/api/core/bitstreams/0d7e8935-879e-4ab6-8c07-e41f7be1465b/content","editor":"AIBD"}]}},"provenance":{"kind":"legacy_record","basis":"Editorial assignment.","record_fields":["date_observed","date_basis","date_precision","revisions"],"source_refs":[]},"review_state":"record_mapped","coverage":"partial"},"replication.attempts[]":{"state":"unresolved","provenance":{"kind":"curatorial","basis":"Editorial assignment.","record_fields":[],"source_refs":[]},"reason":"This is an explicit unresolved assignment: the sources on record do not settle this question.","review_state":"not_assessed","coverage":"partial"},"replication.conditions":{"state":"value_present","value":{"kind":"record_fields","fields":{"replication":{"produced_by":["Each product was set to a threshold giving 90% sensitivity against microbiological results.","The image libraries came from earlier screening studies and were never shared with the makers."],"must_hold":["Programs calibrate the threshold for their own setting, hardware and population.","Programs have laboratory capacity for the confirmatory tests that positive screens require."],"echoes":[]}}},"provenance":{"kind":"legacy_record","basis":"Editorial assignment.","record_fields":["replication"],"source_refs":[]},"review_state":"record_mapped","coverage":"partial"},"revision_history[]":{"state":"value_present","value":{"kind":"record_fields","fields":{"revisions":[{"date":"2026-10-31","change":"Added the record from the finding of WHO.","source":"https://iris.who.int/server/api/core/bitstreams/0d7e8935-879e-4ab6-8c07-e41f7be1465b/content","editor":"AIBD"}]}},"provenance":{"kind":"legacy_record","basis":"Editorial assignment.","record_fields":["revisions"],"source_refs":[]},"review_state":"record_mapped","coverage":"partial"},"sector":{"state":"value_present","value":{"kind":"record_fields","fields":{"domain":"Healthcare (tuberculosis screening)"}},"provenance":{"kind":"legacy_record","basis":"Editorial assignment.","record_fields":["domain"],"source_refs":[]},"review_state":"record_mapped","coverage":"partial"},"sources[].access_version":{"state":"unresolved","provenance":{"kind":"curatorial","basis":"Editorial assignment.","record_fields":[],"source_refs":[]},"reason":"This is an explicit unresolved assignment: the sources on record do not settle this question.","review_state":"not_assessed","coverage":"partial"},"sources[].citation_locator":{"state":"value_present","value":{"kind":"record_fields","fields":{"evidence":[{"kind":"assessment report","citation":"World Health Organization, Global Programme on Tuberculosis & Lung Health. Use of computer-aided detection software for tuberculosis screening: WHO policy statement. Geneva, 10 June 2025. ISBN 978-92-4-011037-3.","title":"Use of computer-aided detection software for tuberculosis screening: WHO policy statement","url":"https://iris.who.int/server/api/core/bitstreams/0d7e8935-879e-4ab6-8c07-e41f7be1465b/content","date":"2025-06-10","written_by":"independent","quote":"CAD4TB v7, DrAid for TB screening v2.4.4–6, Genki Edge v3.4.2, InferRead DR Chest v1.0.1.1, Insight CXR v3.1 and qXR v4.0 met the WHO performance standards for the use of CAD in community-based and facility-based screening for pulmonary TB in people aged 15 years and older"}]}},"provenance":{"kind":"legacy_record","basis":"Editorial assignment.","record_fields":["evidence"],"source_refs":[]},"review_state":"record_mapped","coverage":"partial"},"sources[].id":{"state":"value_present","value":{"kind":"identifiers","identifiers":["AIBD-0733-source-0001"],"role":"Existing evidence items, in their original order"},"provenance":{"kind":"curatorial","basis":"Editorial assignment.","record_fields":[],"source_refs":[]},"review_state":"record_mapped","coverage":"complete"},"sources[].language":{"state":"unresolved","provenance":{"kind":"curatorial","basis":"Editorial assignment.","record_fields":[],"source_refs":[]},"reason":"This is an explicit unresolved assignment: the sources on record do not settle this question.","review_state":"not_assessed","coverage":"partial"},"sources[].relationship":{"state":"value_present","value":{"kind":"record_fields","fields":{"evidence":[{"kind":"assessment report","citation":"World Health Organization, Global Programme on Tuberculosis & Lung Health. Use of computer-aided detection software for tuberculosis screening: WHO policy statement. Geneva, 10 June 2025. ISBN 978-92-4-011037-3.","title":"Use of computer-aided detection software for tuberculosis screening: WHO policy statement","url":"https://iris.who.int/server/api/core/bitstreams/0d7e8935-879e-4ab6-8c07-e41f7be1465b/content","date":"2025-06-10","written_by":"independent","quote":"CAD4TB v7, DrAid for TB screening v2.4.4–6, Genki Edge v3.4.2, InferRead DR Chest v1.0.1.1, Insight CXR v3.1 and qXR v4.0 met the WHO performance standards for the use of CAD in community-based and facility-based screening for pulmonary TB in people aged 15 years and older"}]}},"provenance":{"kind":"legacy_record","basis":"Editorial assignment.","record_fields":["evidence"],"source_refs":[]},"review_state":"record_mapped","coverage":"partial"},"sources[].upstream_sources":{"state":"unresolved","provenance":{"kind":"curatorial","basis":"Editorial assignment.","record_fields":[],"source_refs":[]},"reason":"This is an explicit unresolved assignment: the sources on record do not settle this question.","review_state":"not_assessed","coverage":"partial"},"strategic_assessments[]":{"state":"unresolved","review_state":"not_assessed","coverage":"partial","family":"strategic assessments"},"summary":{"state":"value_present","value":{"kind":"record_fields","fields":{"summary":"WHO found that six of eight AI programs tested on 3,915 archived chest X-rays met its accuracy standards for TB screening of people aged 15 and older.","account":"Computer-aided detection software reads a digital chest X-ray and returns a score for the likelihood of tuberculosis. WHO says that such systems are trained on thousands or millions of images using artificial intelligence. In 2024, WHO invited makers to submit products for an independent evaluation on a validation platform run by FIND in Geneva. Eight products met regulatory requirements. They were tested on two image libraries, 1,663 images for community screening and 2,252 for facility screening, which the makers had not used for training. Microbiological test results served as the reference, and radiologists read the same images for comparison. To pass, a product needed at least 40% specificity at 90% sensitivity and a high partial area under the curve in both libraries. WHO's Technical Advisory Group reviewed the results in January 2025. In a policy statement published on 10 June 2025, WHO found that six products met its standards. In community screening their specificity at 90% sensitivity ranged from 45% to 56%. Accuracy was generally lower for people living with HIV, women and people with a history of TB."}},"provenance":{"kind":"legacy_record","basis":"Editorial assignment.","record_fields":["summary","account"],"source_refs":[]},"review_state":"record_mapped","coverage":"partial"},"systems[].autonomy_permissions":{"state":"unresolved","provenance":{"kind":"curatorial","basis":"Editorial assignment.","record_fields":[],"source_refs":[]},"reason":"This is an explicit unresolved assignment: the sources on record do not settle this question.","review_state":"not_assessed","coverage":"partial"},"systems[].component_role":{"state":"unresolved","provenance":{"kind":"curatorial","basis":"Editorial assignment.","record_fields":[],"source_refs":[]},"reason":"This is an explicit unresolved assignment: the sources on record do not settle this question.","review_state":"not_assessed","coverage":"partial"},"systems[].deployment_configuration":{"state":"unresolved","provenance":{"kind":"curatorial","basis":"Editorial assignment.","record_fields":[],"source_refs":[]},"reason":"This is an explicit unresolved assignment: the sources on record do not settle this question.","review_state":"not_assessed","coverage":"partial"},"systems[].inputs_outputs":{"state":"unresolved","provenance":{"kind":"curatorial","basis":"Editorial assignment.","record_fields":[],"source_refs":[]},"reason":"This is an explicit unresolved assignment: the sources on record do not settle this question.","review_state":"not_assessed","coverage":"partial"},"systems[].model_name":{"state":"unresolved","provenance":{"kind":"curatorial","basis":"Editorial assignment.","record_fields":[],"source_refs":[]},"reason":"This is an explicit unresolved assignment: the sources on record do not settle this question.","review_state":"not_assessed","coverage":"partial"},"systems[].model_version":{"state":"value_present","value":{"kind":"record_fields","fields":{"model_version":"CAD4TB v7, DrAid for TB screening v2.4.4–6, Genki Edge v3.4.2, InferRead DR Chest v1.0.1.1, Insight CXR v3.1 and qXR v4.0"}},"provenance":{"kind":"legacy_record","basis":"Editorial assignment.","record_fields":["model_version"],"source_refs":[]},"review_state":"record_mapped","coverage":"partial"},"systems[].product_name":{"state":"value_present","value":{"kind":"record_fields","fields":{"system":"CAD4TB, DrAid for TB screening, Genki Edge, InferRead DR Chest, Insight CXR and qXR (computer-aided detection software for TB)"}},"provenance":{"kind":"legacy_record","basis":"Editorial assignment.","record_fields":["system"],"source_refs":[]},"review_state":"record_mapped","coverage":"partial"},"systems[].provider":{"state":"unresolved","provenance":{"kind":"curatorial","basis":"Editorial assignment.","record_fields":[],"source_refs":[]},"reason":"This is an explicit unresolved assignment: the sources on record do not settle this question.","review_state":"not_assessed","coverage":"partial"},"taxonomy_crosswalks[]":{"state":"unresolved","provenance":{"kind":"curatorial","basis":"Editorial assignment.","record_fields":[],"source_refs":[]},"reason":"This is an explicit unresolved assignment: the sources on record do not settle this question.","review_state":"not_assessed","coverage":"partial"},"title":{"state":"value_present","value":{"kind":"record_fields","fields":{"title":"WHO finds that six AI chest X-ray programs, including qXR v4.0 and CAD4TB v7, meet its standards for TB screening of people aged 15 and older","headline":"Six AI chest X-ray readers meet WHO's standard for TB screening"}},"provenance":{"kind":"legacy_record","basis":"Editorial assignment.","record_fields":["title","headline"],"source_refs":[]},"review_state":"record_mapped","coverage":"partial"},"version":{"state":"value_present","value":{"kind":"identifiers","identifiers":["3c11ae882c6382c087a3c89dc7220224ff0cbade7f0a4a4b51db67487ea03109"],"role":"SHA-256 of the complete public record and its assessments"},"provenance":{"kind":"curatorial","basis":"SHA-256 of the public record, its assessments, and its displayed answers and notes, excluding its own identifier and the derived coverage summary.","record_fields":[],"source_refs":[]},"review_state":"record_mapped","coverage":"complete"},"workflow":{"state":"value_present","value":{"kind":"record_fields","fields":{"account":"Computer-aided detection software reads a digital chest X-ray and returns a score for the likelihood of tuberculosis. WHO says that such systems are trained on thousands or millions of images using artificial intelligence. In 2024, WHO invited makers to submit products for an independent evaluation on a validation platform run by FIND in Geneva. Eight products met regulatory requirements. They were tested on two image libraries, 1,663 images for community screening and 2,252 for facility screening, which the makers had not used for training. Microbiological test results served as the reference, and radiologists read the same images for comparison. To pass, a product needed at least 40% specificity at 90% sensitivity and a high partial area under the curve in both libraries. WHO's Technical Advisory Group reviewed the results in January 2025. In a policy statement published on 10 June 2025, WHO found that six products met its standards. In community screening their specificity at 90% sensitivity ranged from 45% to 56%. Accuracy was generally lower for people living with HIV, women and people with a history of TB.","replication":{"produced_by":["Each product was set to a threshold giving 90% sensitivity against microbiological results.","The image libraries came from earlier screening studies and were never shared with the makers."],"must_hold":["Programs calibrate the threshold for their own setting, hardware and population.","Programs have laboratory capacity for the confirmatory tests that positive screens require."],"echoes":[]}}},"provenance":{"kind":"legacy_record","basis":"Editorial assignment.","record_fields":["account","replication"],"source_refs":[]},"review_state":"record_mapped","coverage":"partial"}},"summary":{"total":91,"addressed":91,"known":29,"reviewed":2,"reviewed_complete":2,"reviewed_partial":0,"record_mapped":27,"review_not_assessed":62,"not_assessed":0,"complete":false,"by_state":{"value_present":29,"not_reported":0,"unresolved":62,"not_assessed":0,"not_applicable":0,"withheld":0,"source_unavailable":0}}},"presentation_version":"10d6768da7196a067f87353a428b7bb43ff92d621fb040499630b99da436a273"}