  "id": "AIBD-0028",
  "title": "Medtronic MiniMed predictive low-glucose suspend reducing hypoglycemia with a successful pregnancy",
  "summary": "MiniMed 640G predictive low-glucose suspend cut hypoglycemia in a T1D woman after two miscarriages; term birth of a 4,137 g boy.",
  "date_observed": "2021-03-06",
  "date_basis": "unassessed",
  "date_precision": "unassessed",
  "benefit_status": null,
  "source_collection": null,
  "source_entry": null,
  "account": "A case report appeared in Diabetology International on 6 March 2021 (event dates not reported). It describes a 22-year-old Brazilian woman in Japan with type 1 diabetes for 11 years. She miscarried at 7 and 9 weeks in two earlier pregnancies. At about 6 weeks of her third pregnancy she was switched to a Medtronic MiniMed 640G sensor-augmented pump with predictive low glucose suspend. The pump suspended insulin for about 249 minutes per day. Hypoglycemic episodes fell from 2.1–2.5 to 1.4 per day and time below 70 mg/dL from 1.1–1.4% to 0.5%. She delivered a 4,137 g boy by cesarean section at 39 weeks and 3 days with no significant complications.",
  "category": "benefit created",
  "unexpected": false,
  "purpose": "Predictive low glucose suspend is built to stop insulin delivery when sensor glucose is predicted to fall low, to avoid hypoglycemia. The benefit was that purpose, with a successful pregnancy as the downstream outcome.",
  "system": "Medtronic MiniMed sensor-augmented pump / automated insulin delivery with predictive low glucose suspend (SmartGuard)",
  "model_version": "MiniMed 640G (event); MiniMed 780G in echo",
  "developer": "Medtronic",
  "deployer": null,
  "location": null,
  "domain": "Healthcare (diabetes; obstetrics)",
  "segment": "Diabetes technology",
  "benefit_type": "Improved Physical Health",
  "taxonomy_note": null,
  "registry_ids": {},
  "ratings": [
    {
      "group": "personal",
      "depth": 3,
      "reason": "After two miscarriages, hypoglycemia frequency and glucose variability fell on the device and the pregnancy reached term with a healthy infant. A life-changing benefit, though the source does not prove the device caused the live birth."
    }
  ],
  "ai_role": {
    "status": "assessed",
    "statement": "The pump's predictive algorithm autonomously suspended insulin for about four hours a day. The authors link this to fewer hypoglycemic episodes and lower variability. Obstetric and diabetes care were delivered by clinicians. The causal link to carrying the pregnancy is inferred, not demonstrated."
  },
  "evidence": [
    {
      "kind": "case report",
      "citation": "Honoki H, Yagi K, Tsuda S, Wada T, Enkaku A, Nakagawa-Yokoyama M, Kamigishi M, Shikata M, Takikawa-Nishida A, Fujisaka S, Chujo D, Sasaoka T, Nakashima A, Saito S, Tobe K (2021). Type 1 diabetes woman with repeated miscarriages successfully gave birth after introducing an insulin pump with a predictive low glucose suspend feature. Diabetology International 12:324-329.",
      "title": "Type 1 diabetes woman with repeated miscarriages successfully gave birth after introducing an insulin pump with a predictive low glucose suspend feature",
      "url": "https://doi.org/10.1007/s13340-020-00489-9",
      "date": "2021-03-06",
      "written_by": "deployer",
      "quote": "She gave birth to a 4137 g boy at 39 weeks without significant complications."
    },
    {
      "kind": "case report",
      "citation": "Honoki H et al. (2021). Diabetology International 12:324-329.",
      "title": "Type 1 diabetes woman with repeated miscarriages successfully gave birth after introducing an insulin pump with a predictive low glucose suspend feature",
      "url": "https://pmc.ncbi.nlm.nih.gov/articles/PMC8172753/",
      "date": "2021-03-06",
      "written_by": "deployer",
      "quote": "SAP-PLGS makes it possible to avoid severe hypoglycemia following the compensation of insulin to mitigate transient hyperglycemia."
    },
    {
      "kind": "case report",
      "citation": "Zadáková E, Horová E (2025). Case Report: Hybrid closed-loop insulin pump can significantly improve awareness of hypoglycemia. Frontiers in Clinical Diabetes and Healthcare 6:1536038.",
      "title": "Hybrid closed-loop insulin pump can significantly improve awareness of hypoglycemia",
      "url": "https://europepmc.org/article/PMC/PMC11886589",
      "date": "2025-02-21",
      "written_by": "deployer",
      "quote": "since the initiation of the AID, there has been no episode of severe hypoglycemia"
    }
  ],
  "replication": {
    "produced_by": [
      "A person with type 1 diabetes and frequent or unrecognized hypoglycemia uses a Medtronic pump with CGM and predictive low glucose suspend enabled",
      "Sensor glucose is predicted to fall and the algorithm suspends basal insulin",
      "Hypoglycemia frequency and variability fall and the clinical goal (pregnancy, restored awareness) is reached"
    ],
    "must_hold": [
      "Sensor is worn and calibrated with suspend enabled",
      "The user does not override suspensions",
      "Clinical follow-up adjusts settings"
    ],
    "echoes": [
      {
        "date": "2025-02-21",
        "setting": "45-year-old man with type 1 diabetes since 1996 and impaired hypoglycemia awareness, Czechia; MiniMed 780G SmartGuard started mid-2021 after his driving license was suspended in April 2021; no severe hypoglycemia since, Clarke and Gold scores 6 to 2 by early 2022, license reinstated, time below range 23% to 1% by June 2024 (publication date)",
        "model_version": "Medtronic MiniMed 780G (SmartGuard)",
        "outcome": "recurred",
        "source": "Zadáková & Horová (2025). Front Clin Diabetes Healthc 6:1536038. https://pmc.ncbi.nlm.nih.gov/articles/PMC11886589/"
      }
    ]
  },
  "paired_incident": null,
  "headline": "MiniMed 640G cuts hypoglycemia; mother delivers a 4,137 g boy",
  "metric": "hypoglycemic episodes fell from 2.1–2.5 to 1.4 per day",
  "comparator": {
    "against": "her two earlier pregnancies on a MiniMed 620G sensor-augmented pump without predictive low glucose suspend",
    "before": "Hypoglycemic episodes (/day) 2.5 and 2.1; time with AUC below 70 mg/dL 1.1% and 1.4%",
    "after": "Hypoglycemic episodes (/day) 1.4; mean AUC < 70 mg/dL of 0.5%",
    "basis": "measured",
    "quote": "Ratio of time range with average AUC blow 70 mg/dL (%) 1.1 1.4 0.5 Hypoglycemic episodes (/day) 2.5 2.1 1.4",
    "url": "https://pmc.ncbi.nlm.nih.gov/articles/PMC8172753/"
  },
  "later": {
    "status": "unknown",
    "date": null,
    "note": "No development for this patient after the 2021 case report was found; the 2025 MiniMed 780G report concerns a different patient.",
    "quote": null,
    "url": null
  },
  "value": null,
  "developer_country": "Ireland",
  "deployer_country": null,
  "unrated_because": null,
  "brief_notes": null,
  "brief": null,
  "revisions": [
    {
      "date": "2026-10-20",
      "change": "Record drafted from the cited sources.",
      "source": null,
      "editor": "AIBD"
    },
    {
      "date": "2026-10-21",
      "change": "Benefit type set to Improved Physical Health under Trustible's full taxonomy of 23 types; it was Improved Accessibility among the six first used.",
      "source": "https://trustible.ai/post/a-framework-for-measuring-the-benefits-of-ai/",
      "editor": "AIBD"
    },
    {
      "date": "2026-10-21",
      "change": "Country of the developer added from the organization's own site or a registry.",
      "source": "https://en.wikipedia.org/wiki/Medtronic",
      "editor": "AIBD"
    },
    {
      "date": "2026-10-21",
      "change": "Prose pass: a rating reason, account, purpose, the replication conditions, the system's part rewritten in shorter sentences with abbreviations expanded; facts, figures and quotations unchanged.",
      "source": null,
      "editor": "AIBD"
    }
  ],
  "confidence": "claimed",
  "shade": "low",
  "claim_status": "The deployer's account",
  "claim_basis": "Sources: three case reports from the deployer. The cited evidence includes no independent source and no finding by a regulator or court.",
  "profile": {
    "case_id": "AIBD-0028",
    "catalog_version": "1.0.0",
    "source_fingerprint": "bc612c5c94bfaded6e8e94f6847b0dc67055260e052df115570b962bac88c0b1",
    "collection_refs": [
      {
        "collection_id": "aibd",
        "version": "snapshot-8bf7cfec9cb185aa"
      },
      {
        "collection_id": "aibd",
        "version": "2026-10-25"
      }
    ],
    "source_ids": [
      "AIBD-0028-source-0001",
      "AIBD-0028-source-0002",
      "AIBD-0028-source-0003"
    ],
    "families": {
      "strategic assessments": {
        "state": "not_assessed",
        "review_state": "not_assessed",
        "coverage": "partial",
        "record_ids": []
      },
      "evaluations": {
        "state": "not_assessed",
        "review_state": "not_assessed",
        "coverage": "partial",
        "record_ids": []
      },
      "monetary observations": {
        "state": "not_assessed",
        "review_state": "not_assessed",
        "coverage": "partial",
        "record_ids": []
      },
      "institutional decisions": {
        "state": "not_assessed",
        "review_state": "not_assessed",
        "coverage": "partial",
        "record_ids": []
      }
    },
    "fields": {
      "actors[]": {
        "state": "value_present",
        "value": {
          "kind": "record_fields",
          "fields": {
            "developer": "Medtronic"
          }
        },
        "provenance": {
          "kind": "legacy_record",
          "basis": "Preserved from the existing AIBD record; external sources were not rechecked.",
          "record_fields": [
            "developer"
          ],
          "source_refs": []
        },
        "review_state": "record_mapped",
        "coverage": "partial"
      },
      "actors[].ownership_control": {
        "state": "not_assessed",
        "review_state": "not_assessed",
        "coverage": "partial"
      },
      "ai_contribution": {
        "state": "value_present",
        "value": {
          "kind": "record_fields",
          "fields": {
            "ai_role": {
              "status": "assessed",
              "statement": "The pump's predictive algorithm autonomously suspended insulin for about four hours a day. The authors link this to fewer hypoglycemic episodes and lower variability. Obstetric and diabetes care were delivered by clinicians. The causal link to carrying the pregnancy is inferred, not demonstrated."
            }
          }
        },
        "provenance": {
          "kind": "legacy_record",
          "basis": "Preserved from the existing AIBD record; external sources were not rechecked.",
          "record_fields": [
            "ai_role"
          ],
          "source_refs": []
        },
        "review_state": "record_mapped",
        "coverage": "partial"
      },
      "ai_relationship": {
        "state": "value_present",
        "value": {
          "kind": "record_fields",
          "fields": {
            "category": "benefit created",
            "ai_role": {
              "status": "assessed",
              "statement": "The pump's predictive algorithm autonomously suspended insulin for about four hours a day. The authors link this to fewer hypoglycemic episodes and lower variability. Obstetric and diabetes care were delivered by clinicians. The causal link to carrying the pregnancy is inferred, not demonstrated."
            }
          }
        },
        "provenance": {
          "kind": "legacy_record",
          "basis": "Preserved from the existing AIBD record; external sources were not rechecked.",
          "record_fields": [
            "ai_role",
            "category"
          ],
          "source_refs": []
        },
        "review_state": "record_mapped",
        "coverage": "partial"
      },
      "assessments[].alternatives": {
        "state": "not_assessed",
        "review_state": "not_assessed",
        "coverage": "partial"
      },
      "assessments[].assessor_date": {
        "state": "not_assessed",
        "review_state": "not_assessed",
        "coverage": "partial"
      },
      "assessments[].depth": {
        "state": "value_present",
        "value": {
          "kind": "record_fields",
          "fields": {
            "ratings": [
              {
                "group": "personal",
                "depth": 3,
                "reason": "After two miscarriages, hypoglycemia frequency and glucose variability fell on the device and the pregnancy reached term with a healthy infant. A life-changing benefit, though the source does not prove the device caused the live birth."
              }
            ]
          }
        },
        "provenance": {
          "kind": "legacy_record",
          "basis": "Preserved from the existing AIBD record; external sources were not rechecked.",
          "record_fields": [
            "ratings"
          ],
          "source_refs": []
        },
        "review_state": "record_mapped",
        "coverage": "partial"
      },
      "assessments[].display_confidence": {
        "state": "value_present",
        "value": {
          "kind": "structured",
          "data": {
            "confidence": "low",
            "evidence_word": "claimed",
            "claim_status": "The deployer's account",
            "basis": "Sources: three case reports from the deployer. The cited evidence includes no independent source and no finding by a regulator or court.",
            "interpretation": "Existing display rule preserved; the richer outcome-specific confidence question still needs review."
          }
        },
        "provenance": {
          "kind": "curatorial",
          "basis": "Derived from the existing case display properties; not a new evidentiary judgment.",
          "record_fields": [],
          "source_refs": []
        },
        "review_state": "record_mapped",
        "coverage": "partial"
      },
      "assessments[].rationale": {
        "state": "value_present",
        "value": {
          "kind": "record_fields",
          "fields": {
            "ratings": [
              {
                "group": "personal",
                "depth": 3,
                "reason": "After two miscarriages, hypoglycemia frequency and glucose variability fell on the device and the pregnancy reached term with a healthy infant. A life-changing benefit, though the source does not prove the device caused the live birth."
              }
            ]
          }
        },
        "provenance": {
          "kind": "legacy_record",
          "basis": "Preserved from the existing AIBD record; external sources were not rechecked.",
          "record_fields": [
            "ratings"
          ],
          "source_refs": []
        },
        "review_state": "record_mapped",
        "coverage": "partial"
      },
      "assessments[].rubric_version": {
        "state": "not_assessed",
        "review_state": "not_assessed",
        "coverage": "partial"
      },
      "assessments[].status_basis": {
        "state": "not_assessed",
        "review_state": "not_assessed",
        "coverage": "partial"
      },
      "assessments[].target": {
        "state": "value_present",
        "value": {
          "kind": "record_fields",
          "fields": {
            "ratings": [
              {
                "group": "personal",
                "depth": 3,
                "reason": "After two miscarriages, hypoglycemia frequency and glucose variability fell on the device and the pregnancy reached term with a healthy infant. A life-changing benefit, though the source does not prove the device caused the live birth."
              }
            ]
          }
        },
        "provenance": {
          "kind": "legacy_record",
          "basis": "Preserved from the existing AIBD record; external sources were not rechecked.",
          "record_fields": [
            "ratings"
          ],
          "source_refs": []
        },
        "review_state": "record_mapped",
        "coverage": "partial"
      },
      "availability": {
        "state": "not_assessed",
        "review_state": "not_assessed",
        "coverage": "partial"
      },
      "brief": {
        "state": "not_assessed",
        "review_state": "not_assessed",
        "coverage": "partial"
      },
      "case_kind": {
        "state": "not_assessed",
        "review_state": "not_assessed",
        "coverage": "partial"
      },
      "case_links[]": {
        "state": "not_assessed",
        "review_state": "not_assessed",
        "coverage": "partial"
      },
      "claims[].evidence_links": {
        "state": "not_assessed",
        "review_state": "not_assessed",
        "coverage": "partial"
      },
      "claims[].evidence_strength": {
        "state": "not_assessed",
        "review_state": "not_assessed",
        "coverage": "partial"
      },
      "claims[].missingness": {
        "state": "not_assessed",
        "review_state": "not_assessed",
        "coverage": "partial"
      },
      "claims[].proposition": {
        "state": "not_assessed",
        "review_state": "not_assessed",
        "coverage": "partial"
      },
      "cross_border_flows[]": {
        "state": "not_assessed",
        "review_state": "not_assessed",
        "coverage": "partial"
      },
      "decisions[].action_outcome": {
        "state": "not_assessed",
        "review_state": "not_assessed",
        "coverage": "partial",
        "family": "institutional decisions"
      },
      "decisions[].grounds": {
        "state": "not_assessed",
        "review_state": "not_assessed",
        "coverage": "partial",
        "family": "institutional decisions"
      },
      "decisions[].institution_task": {
        "state": "not_assessed",
        "review_state": "not_assessed",
        "coverage": "partial",
        "family": "institutional decisions"
      },
      "decisions[].success_criterion": {
        "state": "not_assessed",
        "review_state": "not_assessed",
        "coverage": "partial",
        "family": "institutional decisions"
      },
      "deployment_stage": {
        "state": "not_assessed",
        "review_state": "not_assessed",
        "coverage": "partial"
      },
      "economics[].amount": {
        "state": "not_assessed",
        "review_state": "not_assessed",
        "coverage": "partial",
        "family": "monetary observations"
      },
      "economics[].costs[]": {
        "state": "not_assessed",
        "review_state": "not_assessed",
        "coverage": "partial",
        "family": "monetary observations"
      },
      "economics[].counterfactual": {
        "state": "not_assessed",
        "review_state": "not_assessed",
        "coverage": "partial",
        "family": "monetary observations"
      },
      "economics[].currency_price_basis": {
        "state": "not_assessed",
        "review_state": "not_assessed",
        "coverage": "partial",
        "family": "monetary observations"
      },
      "economics[].gross_net": {
        "state": "not_assessed",
        "review_state": "not_assessed",
        "coverage": "partial",
        "family": "monetary observations"
      },
      "economics[].measure_type": {
        "state": "not_assessed",
        "review_state": "not_assessed",
        "coverage": "partial",
        "family": "monetary observations"
      },
      "economics[].period_unit": {
        "state": "not_assessed",
        "review_state": "not_assessed",
        "coverage": "partial",
        "family": "monetary observations"
      },
      "economics[].realization": {
        "state": "not_assessed",
        "review_state": "not_assessed",
        "coverage": "partial",
        "family": "monetary observations"
      },
      "economics[].recipient_payer": {
        "state": "not_assessed",
        "review_state": "not_assessed",
        "coverage": "partial",
        "family": "monetary observations"
      },
      "economics[].return_payback": {
        "state": "not_assessed",
        "review_state": "not_assessed",
        "coverage": "partial",
        "family": "monetary observations"
      },
      "economics[].valuation_method": {
        "state": "not_assessed",
        "review_state": "not_assessed",
        "coverage": "partial",
        "family": "monetary observations"
      },
      "evaluations[].ai_uplift": {
        "state": "not_assessed",
        "review_state": "not_assessed",
        "coverage": "partial",
        "family": "evaluations"
      },
      "evaluations[].comparator": {
        "state": "not_assessed",
        "review_state": "not_assessed",
        "coverage": "partial",
        "family": "evaluations"
      },
      "evaluations[].design": {
        "state": "not_assessed",
        "review_state": "not_assessed",
        "coverage": "partial",
        "family": "evaluations"
      },
      "evaluations[].effect": {
        "state": "not_assessed",
        "review_state": "not_assessed",
        "coverage": "partial",
        "family": "evaluations"
      },
      "evaluations[].exposure": {
        "state": "not_assessed",
        "review_state": "not_assessed",
        "coverage": "partial",
        "family": "evaluations"
      },
      "evaluations[].id": {
        "state": "not_assessed",
        "review_state": "not_assessed",
        "coverage": "partial",
        "family": "evaluations"
      },
      "evaluations[].measurement_method": {
        "state": "not_assessed",
        "review_state": "not_assessed",
        "coverage": "partial",
        "family": "evaluations"
      },
      "evaluations[].metric": {
        "state": "not_assessed",
        "provenance": {
          "kind": "legacy_record",
          "basis": "Mapped from the existing case description without a new external source review.",
          "record_fields": [
            "metric"
          ],
          "source_refs": []
        },
        "review_state": "not_assessed",
        "coverage": "partial",
        "family": "evaluations",
        "display": {
          "answer": "hypoglycemic episodes fell from 2.1–2.5 to 1.4 per day",
          "editor_note": ""
        }
      },
      "evaluations[].reproduction_materials": {
        "state": "not_assessed",
        "review_state": "not_assessed",
        "coverage": "partial",
        "family": "evaluations"
      },
      "evaluations[].setting_sample": {
        "state": "not_assessed",
        "review_state": "not_assessed",
        "coverage": "partial",
        "family": "evaluations"
      },
      "evaluations[].uncertainty_limitations": {
        "state": "not_assessed",
        "review_state": "not_assessed",
        "coverage": "partial",
        "family": "evaluations"
      },
      "external_references[]": {
        "state": "not_assessed",
        "review_state": "not_assessed",
        "coverage": "partial"
      },
      "follow_up[]": {
        "state": "value_present",
        "value": {
          "kind": "record_fields",
          "fields": {
            "later": {
              "status": "unknown",
              "note": "No development for this patient after the 2021 case report was found; the 2025 MiniMed 780G report concerns a different patient."
            }
          }
        },
        "provenance": {
          "kind": "legacy_record",
          "basis": "Preserved from the existing AIBD record; external sources were not rechecked.",
          "record_fields": [
            "later"
          ],
          "source_refs": []
        },
        "review_state": "record_mapped",
        "coverage": "partial"
      },
      "human_oversight": {
        "state": "not_assessed",
        "review_state": "not_assessed",
        "coverage": "partial"
      },
      "id": {
        "state": "value_present",
        "value": {
          "kind": "record_fields",
          "fields": {
            "id": "AIBD-0028"
          }
        },
        "provenance": {
          "kind": "legacy_record",
          "basis": "Preserved from the existing AIBD record; external sources were not rechecked.",
          "record_fields": [
            "id"
          ],
          "source_refs": []
        },
        "review_state": "record_mapped",
        "coverage": "partial"
      },
      "infrastructure_dependencies[]": {
        "state": "not_assessed",
        "review_state": "not_assessed",
        "coverage": "partial"
      },
      "outcomes[].beneficiaries": {
        "state": "not_assessed",
        "review_state": "not_assessed",
        "coverage": "partial"
      },
      "outcomes[].beneficiary_count": {
        "state": "not_assessed",
        "review_state": "not_assessed",
        "coverage": "partial"
      },
      "outcomes[].benefit_type": {
        "state": "value_present",
        "value": {
          "kind": "record_fields",
          "fields": {
            "benefit_type": "Improved Physical Health"
          }
        },
        "provenance": {
          "kind": "legacy_record",
          "basis": "Preserved from the existing AIBD record; external sources were not rechecked.",
          "record_fields": [
            "benefit_type"
          ],
          "source_refs": []
        },
        "review_state": "record_mapped",
        "coverage": "partial"
      },
      "outcomes[].claim": {
        "state": "value_present",
        "value": {
          "kind": "record_fields",
          "fields": {
            "account": "A case report appeared in Diabetology International on 6 March 2021 (event dates not reported). It describes a 22-year-old Brazilian woman in Japan with type 1 diabetes for 11 years. She miscarried at 7 and 9 weeks in two earlier pregnancies. At about 6 weeks of her third pregnancy she was switched to a Medtronic MiniMed 640G sensor-augmented pump with predictive low glucose suspend. The pump suspended insulin for about 249 minutes per day. Hypoglycemic episodes fell from 2.1–2.5 to 1.4 per day and time below 70 mg/dL from 1.1–1.4% to 0.5%. She delivered a 4,137 g boy by cesarean section at 39 weeks and 3 days with no significant complications.",
            "benefit_type": "Improved Physical Health"
          }
        },
        "provenance": {
          "kind": "legacy_record",
          "basis": "Preserved from the existing AIBD record; external sources were not rechecked.",
          "record_fields": [
            "account",
            "benefit_type"
          ],
          "source_refs": []
        },
        "review_state": "record_mapped",
        "coverage": "partial"
      },
      "outcomes[].distribution": {
        "state": "not_assessed",
        "review_state": "not_assessed",
        "coverage": "partial"
      },
      "outcomes[].duration": {
        "state": "not_assessed",
        "review_state": "not_assessed",
        "coverage": "partial"
      },
      "outcomes[].id": {
        "state": "not_assessed",
        "review_state": "not_assessed",
        "coverage": "partial"
      },
      "outcomes[].stakeholder_group": {
        "state": "value_present",
        "value": {
          "kind": "record_fields",
          "fields": {
            "ratings": [
              {
                "group": "personal",
                "depth": 3,
                "reason": "After two miscarriages, hypoglycemia frequency and glucose variability fell on the device and the pregnancy reached term with a healthy infant. A life-changing benefit, though the source does not prove the device caused the live birth."
              }
            ]
          }
        },
        "provenance": {
          "kind": "legacy_record",
          "basis": "Preserved from the existing AIBD record; external sources were not rechecked.",
          "record_fields": [
            "ratings"
          ],
          "source_refs": []
        },
        "review_state": "record_mapped",
        "coverage": "partial"
      },
      "outcomes[].status": {
        "state": "not_assessed",
        "review_state": "not_assessed",
        "coverage": "partial"
      },
      "outcomes[].tradeoffs": {
        "state": "not_assessed",
        "review_state": "not_assessed",
        "coverage": "partial"
      },
      "outcomes[].unexpected": {
        "state": "value_present",
        "value": {
          "kind": "record_fields",
          "fields": {
            "unexpected": false,
            "purpose": "Predictive low glucose suspend is built to stop insulin delivery when sensor glucose is predicted to fall low, to avoid hypoglycemia. The benefit was that purpose, with a successful pregnancy as the downstream outcome."
          }
        },
        "provenance": {
          "kind": "legacy_record",
          "basis": "Preserved from the existing AIBD record; external sources were not rechecked.",
          "record_fields": [
            "unexpected",
            "purpose"
          ],
          "source_refs": []
        },
        "review_state": "record_mapped",
        "coverage": "partial"
      },
      "places[]": {
        "state": "value_present",
        "value": {
          "kind": "record_fields",
          "fields": {
            "developer_country": "Ireland"
          }
        },
        "provenance": {
          "kind": "legacy_record",
          "basis": "Preserved from the existing AIBD record; external sources were not rechecked.",
          "record_fields": [
            "developer_country"
          ],
          "source_refs": []
        },
        "review_state": "record_mapped",
        "coverage": "partial"
      },
      "purpose": {
        "state": "value_present",
        "value": {
          "kind": "record_fields",
          "fields": {
            "purpose": "Predictive low glucose suspend is built to stop insulin delivery when sensor glucose is predicted to fall low, to avoid hypoglycemia. The benefit was that purpose, with a successful pregnancy as the downstream outcome."
          }
        },
        "provenance": {
          "kind": "legacy_record",
          "basis": "Preserved from the existing AIBD record; external sources were not rechecked.",
          "record_fields": [
            "purpose"
          ],
          "source_refs": []
        },
        "review_state": "record_mapped",
        "coverage": "partial"
      },
      "record_dates": {
        "state": "value_present",
        "value": {
          "kind": "record_fields",
          "fields": {
            "date_observed": "2021-03-06",
            "revisions": [
              {
                "date": "2026-10-20",
                "change": "Record drafted from the cited sources.",
                "editor": "AIBD"
              },
              {
                "date": "2026-10-21",
                "change": "Benefit type set to Improved Physical Health under Trustible's full taxonomy of 23 types; it was Improved Accessibility among the six first used.",
                "source": "https://trustible.ai/post/a-framework-for-measuring-the-benefits-of-ai/",
                "editor": "AIBD"
              },
              {
                "date": "2026-10-21",
                "change": "Country of the developer added from the organization's own site or a registry.",
                "source": "https://en.wikipedia.org/wiki/Medtronic",
                "editor": "AIBD"
              },
              {
                "date": "2026-10-21",
                "change": "Prose pass: a rating reason, account, purpose, the replication conditions, the system's part rewritten in shorter sentences with abbreviations expanded; facts, figures and quotations unchanged.",
                "editor": "AIBD"
              }
            ]
          }
        },
        "provenance": {
          "kind": "legacy_record",
          "basis": "Preserved from the existing AIBD record; external sources were not rechecked.",
          "record_fields": [
            "date_observed",
            "revisions"
          ],
          "source_refs": []
        },
        "review_state": "record_mapped",
        "coverage": "partial"
      },
      "replication.attempts[]": {
        "state": "value_present",
        "value": {
          "kind": "record_fields",
          "fields": {
            "replication": {
              "produced_by": [
                "A person with type 1 diabetes and frequent or unrecognized hypoglycemia uses a Medtronic pump with CGM and predictive low glucose suspend enabled",
                "Sensor glucose is predicted to fall and the algorithm suspends basal insulin",
                "Hypoglycemia frequency and variability fall and the clinical goal (pregnancy, restored awareness) is reached"
              ],
              "must_hold": [
                "Sensor is worn and calibrated with suspend enabled",
                "The user does not override suspensions",
                "Clinical follow-up adjusts settings"
              ],
              "echoes": [
                {
                  "date": "2025-02-21",
                  "setting": "45-year-old man with type 1 diabetes since 1996 and impaired hypoglycemia awareness, Czechia; MiniMed 780G SmartGuard started mid-2021 after his driving license was suspended in April 2021; no severe hypoglycemia since, Clarke and Gold scores 6 to 2 by early 2022, license reinstated, time below range 23% to 1% by June 2024 (publication date)",
                  "model_version": "Medtronic MiniMed 780G (SmartGuard)",
                  "outcome": "recurred",
                  "source": "Zadáková & Horová (2025). Front Clin Diabetes Healthc 6:1536038. https://pmc.ncbi.nlm.nih.gov/articles/PMC11886589/"
                }
              ]
            }
          }
        },
        "provenance": {
          "kind": "legacy_record",
          "basis": "Preserved from the existing AIBD record; external sources were not rechecked.",
          "record_fields": [
            "replication"
          ],
          "source_refs": []
        },
        "review_state": "record_mapped",
        "coverage": "partial"
      },
      "replication.conditions": {
        "state": "value_present",
        "value": {
          "kind": "record_fields",
          "fields": {
            "replication": {
              "produced_by": [
                "A person with type 1 diabetes and frequent or unrecognized hypoglycemia uses a Medtronic pump with CGM and predictive low glucose suspend enabled",
                "Sensor glucose is predicted to fall and the algorithm suspends basal insulin",
                "Hypoglycemia frequency and variability fall and the clinical goal (pregnancy, restored awareness) is reached"
              ],
              "must_hold": [
                "Sensor is worn and calibrated with suspend enabled",
                "The user does not override suspensions",
                "Clinical follow-up adjusts settings"
              ],
              "echoes": [
                {
                  "date": "2025-02-21",
                  "setting": "45-year-old man with type 1 diabetes since 1996 and impaired hypoglycemia awareness, Czechia; MiniMed 780G SmartGuard started mid-2021 after his driving license was suspended in April 2021; no severe hypoglycemia since, Clarke and Gold scores 6 to 2 by early 2022, license reinstated, time below range 23% to 1% by June 2024 (publication date)",
                  "model_version": "Medtronic MiniMed 780G (SmartGuard)",
                  "outcome": "recurred",
                  "source": "Zadáková & Horová (2025). Front Clin Diabetes Healthc 6:1536038. https://pmc.ncbi.nlm.nih.gov/articles/PMC11886589/"
                }
              ]
            }
          }
        },
        "provenance": {
          "kind": "legacy_record",
          "basis": "Preserved from the existing AIBD record; external sources were not rechecked.",
          "record_fields": [
            "replication"
          ],
          "source_refs": []
        },
        "review_state": "record_mapped",
        "coverage": "partial"
      },
      "revision_history[]": {
        "state": "value_present",
        "value": {
          "kind": "record_fields",
          "fields": {
            "revisions": [
              {
                "date": "2026-10-20",
                "change": "Record drafted from the cited sources.",
                "editor": "AIBD"
              },
              {
                "date": "2026-10-21",
                "change": "Benefit type set to Improved Physical Health under Trustible's full taxonomy of 23 types; it was Improved Accessibility among the six first used.",
                "source": "https://trustible.ai/post/a-framework-for-measuring-the-benefits-of-ai/",
                "editor": "AIBD"
              },
              {
                "date": "2026-10-21",
                "change": "Country of the developer added from the organization's own site or a registry.",
                "source": "https://en.wikipedia.org/wiki/Medtronic",
                "editor": "AIBD"
              },
              {
                "date": "2026-10-21",
                "change": "Prose pass: a rating reason, account, purpose, the replication conditions, the system's part rewritten in shorter sentences with abbreviations expanded; facts, figures and quotations unchanged.",
                "editor": "AIBD"
              }
            ]
          }
        },
        "provenance": {
          "kind": "legacy_record",
          "basis": "Preserved from the existing AIBD record; external sources were not rechecked.",
          "record_fields": [
            "revisions"
          ],
          "source_refs": []
        },
        "review_state": "record_mapped",
        "coverage": "partial"
      },
      "sector": {
        "state": "value_present",
        "value": {
          "kind": "record_fields",
          "fields": {
            "domain": "Healthcare (diabetes; obstetrics)",
            "segment": "Diabetes technology"
          }
        },
        "provenance": {
          "kind": "legacy_record",
          "basis": "Preserved from the existing AIBD record; external sources were not rechecked.",
          "record_fields": [
            "domain",
            "segment"
          ],
          "source_refs": []
        },
        "review_state": "record_mapped",
        "coverage": "partial"
      },
      "sources[].access_version": {
        "state": "not_assessed",
        "review_state": "not_assessed",
        "coverage": "partial"
      },
      "sources[].citation_locator": {
        "state": "value_present",
        "value": {
          "kind": "record_fields",
          "fields": {
            "evidence": [
              {
                "kind": "case report",
                "citation": "Honoki H, Yagi K, Tsuda S, Wada T, Enkaku A, Nakagawa-Yokoyama M, Kamigishi M, Shikata M, Takikawa-Nishida A, Fujisaka S, Chujo D, Sasaoka T, Nakashima A, Saito S, Tobe K (2021). Type 1 diabetes woman with repeated miscarriages successfully gave birth after introducing an insulin pump with a predictive low glucose suspend feature. Diabetology International 12:324-329.",
                "title": "Type 1 diabetes woman with repeated miscarriages successfully gave birth after introducing an insulin pump with a predictive low glucose suspend feature",
                "url": "https://doi.org/10.1007/s13340-020-00489-9",
                "date": "2021-03-06",
                "written_by": "deployer",
                "quote": "She gave birth to a 4137 g boy at 39 weeks without significant complications."
              },
              {
                "kind": "case report",
                "citation": "Honoki H et al. (2021). Diabetology International 12:324-329.",
                "title": "Type 1 diabetes woman with repeated miscarriages successfully gave birth after introducing an insulin pump with a predictive low glucose suspend feature",
                "url": "https://pmc.ncbi.nlm.nih.gov/articles/PMC8172753/",
                "date": "2021-03-06",
                "written_by": "deployer",
                "quote": "SAP-PLGS makes it possible to avoid severe hypoglycemia following the compensation of insulin to mitigate transient hyperglycemia."
              },
              {
                "kind": "case report",
                "citation": "Zadáková E, Horová E (2025). Case Report: Hybrid closed-loop insulin pump can significantly improve awareness of hypoglycemia. Frontiers in Clinical Diabetes and Healthcare 6:1536038.",
                "title": "Hybrid closed-loop insulin pump can significantly improve awareness of hypoglycemia",
                "url": "https://europepmc.org/article/PMC/PMC11886589",
                "date": "2025-02-21",
                "written_by": "deployer",
                "quote": "since the initiation of the AID, there has been no episode of severe hypoglycemia"
              }
            ]
          }
        },
        "provenance": {
          "kind": "legacy_record",
          "basis": "Preserved from the existing AIBD record; external sources were not rechecked.",
          "record_fields": [
            "evidence"
          ],
          "source_refs": []
        },
        "review_state": "record_mapped",
        "coverage": "partial"
      },
      "sources[].id": {
        "state": "value_present",
        "value": {
          "kind": "identifiers",
          "identifiers": [
            "AIBD-0028-source-0001",
            "AIBD-0028-source-0002",
            "AIBD-0028-source-0003"
          ],
          "role": "Existing evidence items, in their original order"
        },
        "provenance": {
          "kind": "curatorial",
          "basis": "Stable within the fingerprinted record version; no source content or source review has been added.",
          "record_fields": [],
          "source_refs": []
        },
        "review_state": "record_mapped",
        "coverage": "complete"
      },
      "sources[].language": {
        "state": "not_assessed",
        "review_state": "not_assessed",
        "coverage": "partial"
      },
      "sources[].relationship": {
        "state": "value_present",
        "value": {
          "kind": "record_fields",
          "fields": {
            "evidence": [
              {
                "kind": "case report",
                "citation": "Honoki H, Yagi K, Tsuda S, Wada T, Enkaku A, Nakagawa-Yokoyama M, Kamigishi M, Shikata M, Takikawa-Nishida A, Fujisaka S, Chujo D, Sasaoka T, Nakashima A, Saito S, Tobe K (2021). Type 1 diabetes woman with repeated miscarriages successfully gave birth after introducing an insulin pump with a predictive low glucose suspend feature. Diabetology International 12:324-329.",
                "title": "Type 1 diabetes woman with repeated miscarriages successfully gave birth after introducing an insulin pump with a predictive low glucose suspend feature",
                "url": "https://doi.org/10.1007/s13340-020-00489-9",
                "date": "2021-03-06",
                "written_by": "deployer",
                "quote": "She gave birth to a 4137 g boy at 39 weeks without significant complications."
              },
              {
                "kind": "case report",
                "citation": "Honoki H et al. (2021). Diabetology International 12:324-329.",
                "title": "Type 1 diabetes woman with repeated miscarriages successfully gave birth after introducing an insulin pump with a predictive low glucose suspend feature",
                "url": "https://pmc.ncbi.nlm.nih.gov/articles/PMC8172753/",
                "date": "2021-03-06",
                "written_by": "deployer",
                "quote": "SAP-PLGS makes it possible to avoid severe hypoglycemia following the compensation of insulin to mitigate transient hyperglycemia."
              },
              {
                "kind": "case report",
                "citation": "Zadáková E, Horová E (2025). Case Report: Hybrid closed-loop insulin pump can significantly improve awareness of hypoglycemia. Frontiers in Clinical Diabetes and Healthcare 6:1536038.",
                "title": "Hybrid closed-loop insulin pump can significantly improve awareness of hypoglycemia",
                "url": "https://europepmc.org/article/PMC/PMC11886589",
                "date": "2025-02-21",
                "written_by": "deployer",
                "quote": "since the initiation of the AID, there has been no episode of severe hypoglycemia"
              }
            ]
          }
        },
        "provenance": {
          "kind": "legacy_record",
          "basis": "Preserved from the existing AIBD record; external sources were not rechecked.",
          "record_fields": [
            "evidence"
          ],
          "source_refs": []
        },
        "review_state": "record_mapped",
        "coverage": "partial"
      },
      "sources[].upstream_sources": {
        "state": "not_assessed",
        "review_state": "not_assessed",
        "coverage": "partial"
      },
      "strategic_assessments[]": {
        "state": "not_assessed",
        "review_state": "not_assessed",
        "coverage": "partial",
        "family": "strategic assessments"
      },
      "summary": {
        "state": "value_present",
        "value": {
          "kind": "record_fields",
          "fields": {
            "summary": "MiniMed 640G predictive low-glucose suspend cut hypoglycemia in a T1D woman after two miscarriages; term birth of a 4,137 g boy.",
            "account": "A case report appeared in Diabetology International on 6 March 2021 (event dates not reported). It describes a 22-year-old Brazilian woman in Japan with type 1 diabetes for 11 years. She miscarried at 7 and 9 weeks in two earlier pregnancies. At about 6 weeks of her third pregnancy she was switched to a Medtronic MiniMed 640G sensor-augmented pump with predictive low glucose suspend. The pump suspended insulin for about 249 minutes per day. Hypoglycemic episodes fell from 2.1–2.5 to 1.4 per day and time below 70 mg/dL from 1.1–1.4% to 0.5%. She delivered a 4,137 g boy by cesarean section at 39 weeks and 3 days with no significant complications."
          }
        },
        "provenance": {
          "kind": "legacy_record",
          "basis": "Preserved from the existing AIBD record; external sources were not rechecked.",
          "record_fields": [
            "summary",
            "account"
          ],
          "source_refs": []
        },
        "review_state": "record_mapped",
        "coverage": "partial"
      },
      "systems[].autonomy_permissions": {
        "state": "not_assessed",
        "review_state": "not_assessed",
        "coverage": "partial"
      },
      "systems[].component_role": {
        "state": "value_present",
        "value": {
          "kind": "text",
          "text": "The pump predicted low glucose and suspended insulin delivery."
        },
        "provenance": {
          "kind": "legacy_record",
          "basis": "Restated the mechanical task already described in the existing system and AI-role fields. No new source review is claimed.",
          "record_fields": [
            "system",
            "ai_role"
          ],
          "source_refs": []
        },
        "review_state": "record_mapped",
        "coverage": "partial"
      },
      "systems[].deployment_configuration": {
        "state": "value_present",
        "value": {
          "kind": "text",
          "text": "The initial event involved a MiniMed 640G. A later report involved a MiniMed 780G."
        },
        "provenance": {
          "kind": "legacy_record",
          "basis": "Separated the model, software, hardware, or operation already described in the existing record. No new source review is claimed.",
          "record_fields": [
            "model_version",
            "system",
            "ai_role"
          ],
          "source_refs": []
        },
        "review_state": "record_mapped",
        "coverage": "partial"
      },
      "systems[].inputs_outputs": {
        "state": "not_assessed",
        "review_state": "not_assessed",
        "coverage": "partial"
      },
      "systems[].model_name": {
        "state": "not_assessed",
        "review_state": "not_assessed",
        "coverage": "partial"
      },
      "systems[].model_version": {
        "state": "value_present",
        "value": {
          "kind": "record_fields",
          "fields": {
            "model_version": "MiniMed 640G (event); MiniMed 780G in echo"
          }
        },
        "provenance": {
          "kind": "legacy_record",
          "basis": "Preserved from the existing AIBD record; external sources were not rechecked.",
          "record_fields": [
            "model_version"
          ],
          "source_refs": []
        },
        "review_state": "record_mapped",
        "coverage": "partial",
        "display": {
          "answer": "Exact release not established.",
          "editor_note": ""
        }
      },
      "systems[].product_name": {
        "state": "value_present",
        "value": {
          "kind": "record_fields",
          "fields": {
            "system": "Medtronic MiniMed sensor-augmented pump / automated insulin delivery with predictive low glucose suspend (SmartGuard)"
          }
        },
        "provenance": {
          "kind": "legacy_record",
          "basis": "Preserved from the existing AIBD record; external sources were not rechecked.",
          "record_fields": [
            "system"
          ],
          "source_refs": []
        },
        "review_state": "record_mapped",
        "coverage": "partial"
      },
      "systems[].provider": {
        "state": "value_present",
        "value": {
          "kind": "structured",
          "data": {
            "application_developer": "Medtronic"
          }
        },
        "provenance": {
          "kind": "legacy_record",
          "basis": "Mapped from the existing case description without a new external source review.",
          "record_fields": [
            "developer"
          ],
          "source_refs": []
        },
        "review_state": "record_mapped",
        "coverage": "partial"
      },
      "taxonomy_crosswalks[]": {
        "state": "not_assessed",
        "review_state": "not_assessed",
        "coverage": "partial"
      },
      "title": {
        "state": "value_present",
        "value": {
          "kind": "record_fields",
          "fields": {
            "title": "Medtronic MiniMed predictive low-glucose suspend reducing hypoglycemia with a successful pregnancy",
            "headline": "MiniMed 640G cuts hypoglycemia; mother delivers a 4,137 g boy"
          }
        },
        "provenance": {
          "kind": "legacy_record",
          "basis": "Preserved from the existing AIBD record; external sources were not rechecked.",
          "record_fields": [
            "title",
            "headline"
          ],
          "source_refs": []
        },
        "review_state": "record_mapped",
        "coverage": "partial"
      },
      "version": {
        "state": "value_present",
        "value": {
          "kind": "identifiers",
          "identifiers": [
            "bc612c5c94bfaded6e8e94f6847b0dc67055260e052df115570b962bac88c0b1"
          ],
          "role": "SHA-256 of normalized original case data"
        },
        "provenance": {
          "kind": "curatorial",
          "basis": "Deterministic content version assigned by the local migration.",
          "record_fields": [],
          "source_refs": []
        },
        "review_state": "record_mapped",
        "coverage": "complete"
      },
      "workflow": {
        "state": "value_present",
        "value": {
          "kind": "record_fields",
          "fields": {
            "account": "A case report appeared in Diabetology International on 6 March 2021 (event dates not reported). It describes a 22-year-old Brazilian woman in Japan with type 1 diabetes for 11 years. She miscarried at 7 and 9 weeks in two earlier pregnancies. At about 6 weeks of her third pregnancy she was switched to a Medtronic MiniMed 640G sensor-augmented pump with predictive low glucose suspend. The pump suspended insulin for about 249 minutes per day. Hypoglycemic episodes fell from 2.1–2.5 to 1.4 per day and time below 70 mg/dL from 1.1–1.4% to 0.5%. She delivered a 4,137 g boy by cesarean section at 39 weeks and 3 days with no significant complications.",
            "replication": {
              "produced_by": [
                "A person with type 1 diabetes and frequent or unrecognized hypoglycemia uses a Medtronic pump with CGM and predictive low glucose suspend enabled",
                "Sensor glucose is predicted to fall and the algorithm suspends basal insulin",
                "Hypoglycemia frequency and variability fall and the clinical goal (pregnancy, restored awareness) is reached"
              ],
              "must_hold": [
                "Sensor is worn and calibrated with suspend enabled",
                "The user does not override suspensions",
                "Clinical follow-up adjusts settings"
              ],
              "echoes": [
                {
                  "date": "2025-02-21",
                  "setting": "45-year-old man with type 1 diabetes since 1996 and impaired hypoglycemia awareness, Czechia; MiniMed 780G SmartGuard started mid-2021 after his driving license was suspended in April 2021; no severe hypoglycemia since, Clarke and Gold scores 6 to 2 by early 2022, license reinstated, time below range 23% to 1% by June 2024 (publication date)",
                  "model_version": "Medtronic MiniMed 780G (SmartGuard)",
                  "outcome": "recurred",
                  "source": "Zadáková & Horová (2025). Front Clin Diabetes Healthc 6:1536038. https://pmc.ncbi.nlm.nih.gov/articles/PMC11886589/"
                }
              ]
            }
          }
        },
        "provenance": {
          "kind": "legacy_record",
          "basis": "Preserved from the existing AIBD record; external sources were not rechecked.",
          "record_fields": [
            "account",
            "replication"
          ],
          "source_refs": []
        },
        "review_state": "record_mapped",
        "coverage": "partial"
      }
    },
    "summary": {
      "total": 91,
      "addressed": 91,
      "known": 32,
      "reviewed": 0,
      "record_mapped": 32,
      "not_assessed": 59,
      "complete": false,
      "by_state": {
        "value_present": 32,
        "not_reported": 0,
        "unresolved": 0,
        "not_assessed": 59,
        "not_applicable": 0,
        "withheld": 0,
        "source_unavailable": 0
      }
    }
  }
}